DICOM PS3.3 2024d - Information Object Definitions |
---|
Table C.11.28-1 specifies the Attributes of the Volumetric Graphic Annotation Module, which specify graphic annotation placed in the Volumetric Presentation State Reference Coordinate System coordinate space.
This Module specifies literal text and/or graphic annotation only. Provision is made to reference external measurement Structured Report Instances to provide context for the annotation, if desired.
Table C.11.28-1. Volumetric Graphic Annotation Module Attributes
Graphic annotations described by coordinates in the Volumetric Presentation State Reference Coordinate System. One or more Items shall be included in this Sequence. Required if Volumetric Presentation Input Annotation Sequence (0070,1905) is not present. May be present otherwise. |
|||
An ordered set of (x,y,z) triplets (in mm and may be negative) that define a region of interest in the Volumetric Presentation State Reference Coordinate System defined by Frame of Reference UID (0020,0052). See Section C.18.9.1.1. |
|||
See Section C.18.9.1.2 for Enumerated Values. |
|||
The layer defined in the Section C.10.7 “Graphic Layer Module” in which the graphics or text is to be rendered. |
|||
Specifies whether the volumetric annotation should be displayed only when it intersects with the presentation view. See Section C.11.28.1. |
|||
Displayed text associated with the input Instance. See Section C.10.5.1.1. |
|||
Recommended location of the Top Left Hand Corner (TLHC) of the bounding box in which Unformatted Text Value (0070,0006) is to be placed within a Volumetric Presentation View, specified as fractions of the view width and view height. Each value shall be within the range 0.0 to 1.0. |
|||
Recommended location of the Bottom Right Hand Corner (BRHC) of the bounding box in which Unformatted Text Value (0070,0006) is to be placed within a Volumetric Presentation View, specified as fractions of the view width and view height. Each value shall be within the range 0.0 to 1.0. Required if Bounding Box Top Left Hand Corner (0070,0010) is present. |
|||
Location of the Unformatted Test Value (0070,0006) relative to the vertical edges of the bounding box. Required if Bounding Box Top Left Hand Corner (0070,0010) is present. |
|||
Reference to a node in a Structured Report Instance providing context for this annotation. Only one Item shall be present in this Sequence. See Section C.11.28.2. |
|||
>>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
Reference to a Content Item in the referenced Structured Report specified as the Observation UID (0040,A171) of the Content Item. |
|||
Annotations of inputs in Volumetric Presentation State Input Sequence (0070,1201). One or more Items shall be included in this Sequence. Required if Volumetric Annotation Sequence (0070,1901) is not present. May be present otherwise. |
|||
Value of Volumetric Presentation Input Number (0070,1207) for this input in the Volumetric Presentation State Input Sequence (0070,1201). |
|||
The layer defined in the Section C.10.7 Graphic Layer Module in which the graphic or text is to be rendered. |
|||
Displayed text associated with the input. See Section C.10.5.1.1. |
|||
Recommended location of the Top Left Hand Corner (TLHC) of the bounding box in which Unformatted Text Value (0070,0006) is to be displayed, as fractions of the specified presentation view width and height. Each value shall be within the range 0.0 to 1.0. See Section C.10.5.1.1. |
|||
Recommended location of the Bottom Right Hand Corner (BRHC) of the bounding box in which Unformatted Text Value (0070,0006) is to be displayed, as fractions of the specified presentation view width and height. Each value shall be within the range 0.0 to 1.0. See Section C.10.5.1.1. Required if Bounding Box Top Left Hand Corner (0070,0010) is present. |
|||
Location of the Unformatted Test Value (0070,0006) relative to the vertical edges of the bounding box. See Section C.10.5.1.1. Required if Bounding Box Top Left Hand Corner (0070,0010) is present. |
|||
Reference to a node in a Structured Report Instance providing context for this annotation. Only one Item shall be present in this Sequence. See Section C.11.28.2. |
|||
>>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
Reference to a Content Item in the referenced Structured Report specified as the Observation UID (0040,A171) of the Content Item. |
A graphic defined by the Volumetric Annotation Sequence (0070,1901) may contain data that falls outside the view defined by the Volumetric Presentation State geometry.
If the Volumetric Presentation State intends that an annotation be clipped where it falls outside the defined view, Annotation Clipping (0070,1907) shall have a value of YES.
If the Volumetric Presentation State intends that an annotation be displayed in its entirely even where it does not intersect the volume of interest then Annotation Clipping (0070,1907) shall have a value of NO.
Any visual clues used to indicate where the annotation does and does not intersect the volume of interest are not specified by the Volumetric Presentation State.
DICOM PS3.3 2024d - Information Object Definitions |
---|