DICOM PS3.3 2024d - Information Object Definitions |
---|
Table C.11.19-1 specifies the Attributes of the XA/XRF Presentation State Shutter Module, which describe shutter operations for a XA/XRF Multi-frame image.
Table C.11.20-1. XA/XRF Presentation State Shutter Module Attributes
Sequence of shutter specifications to be applied to groups of frames. |
|||
Sequence of Items where each Item provides reference to a selected set of Image SOP Class/SOP Instance pairs that are defined in the Presentation State Relationship Module. One or more Items shall be included in this Sequence. Required if shutters in this Item do not apply to all the images and frames listed in the Presentation State Relationship Module. |
|||
>>Include Table 10-3 “Image SOP Instance Reference Macro Attributes” |
|||
DICOM PS3.3 2024d - Information Object Definitions |
---|