DICOM PS3.3 2024d - Information Object Definitions

A.75.3 Parametric Map IOD Module Table

Table A.75-1. Parametric Map IOD Modules

IE

Module

Reference

Usage

Patient

Patient

C.7.1.1

M

Clinical Trial Subject

C.7.1.3

U

Study

General Study

C.7.2.1

M

Patient Study

C.7.2.2

U

Clinical Trial Study

C.7.2.3

U

Series

General Series

C.7.3.1

M

Parametric Map Series

C.8.32.1

M

Clinical Trial Series

C.7.3.2

U

Frame of Reference

Frame of Reference

C.7.4.1

M

Synchronization

C.7.4.2

U

Equipment

General Equipment

C.7.5.1

M

Enhanced General Equipment

C.7.5.2

M

Acquisition

General Acquisition

C.7.10.1

U

Multi-Resolution Pyramid

Multi-Resolution Pyramid

C.7.11.1

U

Image

General Image

C.7.6.1

M

General Reference

C.12.4

U

Microscope Slide Layer Tile Organization

C.8.12.14

C - Required if Dimension Organization Type (0020,9311) is present with a value of TILED_FULL. May be present otherwise.

Image Pixel

C.7.6.3

C - Required if integer pixels

Floating Point Image Pixel

C.7.6.24

C - Required if 32 bit floating point pixels

Double Floating Point Image Pixel

C.7.6.25

C - Required if 64 bit floating point pixels

Parametric Map Image

C.8.32.2

M

Multi-frame Functional Groups

C.7.6.16

M

Multi-frame Dimension

C.7.6.17

M

Palette Color Lookup Table

C.7.9

C - Required if Pixel Presentation (0008,9205) in the Parametric Map image Module equals COLOR_RANGE and Palette Color Lookup Table UID (0028,1199) is not present.

Cardiac Synchronization

C.7.6.18.1

U

Respiratory Synchronization

C.7.6.18.2

U

Bulk Motion Synchronization

C.7.6.18.3

U

Acquisition Context

C.7.6.14

M

Device

C.7.6.12

U

Specimen

C.7.6.22

U

Common Instance Reference

C.12.2

C - Required if Referenced Image Functional Group (Section C.7.6.16.2.5) or Derivation Image Functional Group (Section C.7.6.16.2.6) is present.

SOP Common

C.12.1

M

Frame Extraction

C.12.3

C - Required if the SOP Instance was created in response to a Frame-Level retrieve request


DICOM PS3.3 2024d - Information Object Definitions