DICOM PS3.3 2024d - Information Object Definitions

A.5.4 Nuclear Medicine Image IOD Module Table

Table A.5-1 specifies the Modules of the Nuclear Medicine Image IOD.

Table A.5-1. Nuclear Medicine Image IOD Modules

IE

Module

Reference

Usage

Patient

Patient

C.7.1.1

M

Clinical Trial Subject

C.7.1.3

U

Study

General Study

C.7.2.1

M

Patient Study

C.7.2.2

U

Clinical Trial Study

C.7.2.3

U

Series

General Series

C.7.3.1

M

Clinical Trial Series

C.7.3.2

U

NM/PET Patient Orientation

C.8.4.6

M

Frame of Reference

Frame of Reference

C.7.4.1

U

Synchronization

C.7.4.2

C - Required if time synchronization was applied

Equipment

General Equipment

C.7.5.1

M

Acquisition

General Acquisition

C.7.10.1

M

Image

General Image

C.7.6.1

M

General Reference

C.12.4

U

Image Pixel

C.7.6.3

M

Acquisition Context

C.7.6.14

U - See Section A.5.4.1

Device

C.7.6.12

U

Specimen

C.7.6.22

U

NM Image Pixel

C.8.4.7

M

Multi-frame

C.7.6.6

M

NM Multi-frame

C.8.4.8

M

NM Image

C.8.4.9

M

NM Isotope

C.8.4.10

M

NM Detector

C.8.4.11

M

NM Tomo Acquisition

C.8.4.12

C - Required if Image Type (0008,0008) Value 3 is TOMO, GATED TOMO, RECON TOMO or RECON GATED TOMO

NM Multi-gated Acquisition

C.8.4.13

C - Required if Image Type (0008,0008) Value 3 is GATED, GATED TOMO, or RECON GATED TOMO

NM Phase

C.8.4.14

C - Required if Image Type (0008,0008) Value 3 is DYNAMIC

NM Reconstruction

C.8.4.15

C - Required if Image Type (0008,0008) Value 3 is RECON TOMO or RECON GATED TOMO

Overlay Plane

C.9.2

U

Multi-frame Overlay

C.9.3

U

VOI LUT

C.11.2

U

ICC Profile

C.11.15

U

SOP Common

C.12.1

M

Common Instance Reference

C.12.2

U

Frame Extraction

C.12.3

C - Required if the SOP Instance was created in response to a Frame-Level retrieve request


Note

The Curve Module (Retired) was previously included in the Image IE for this IOD but has been retired. See PS3.3-2004.

A.5.4.1 Acquisition Context Module

For Acquisition Context Sequence (0040,0555) DTID 3470 “NM/PET Acquisition Context” shall be used, which includes description of the cardiovascular rest or stress state.

The Acquisition Context Sequence (0040,0555) shall always apply to all frames in the Image. Patient State shall always apply to all frames in the Image, therefore, neither Referenced Frame Numbers (0040,A136) nor Referenced Frame Number (0008,1160) shall be present.

The Acquisition Context information may be entered during acquisition, or obtained from the Modality Worklist using information supplied in the Protocol Context, using TID 15101 “NM/PET Protocol Context”.

DICOM PS3.3 2024d - Information Object Definitions