DICOM PS3.3 2024d - Information Object Definitions

A.48.4 Enhanced XRF Image Functional Group Macros

Table A.48-2 specifies the use of the Functional Group Macros used in the Multi-frame Functional Groups Module for the Enhanced XRF Image IOD.

Table A.48-2. Enhanced XRF Image Functional Group Macros

Functional Group Macro

Section

Usage

Frame Content

C.7.6.16.2.2

M - May not be used as a Shared Functional Group.

Referenced Image

C.7.6.16.2.5

U

Derivation Image

C.7.6.16.2.6

C - Required if the image or frame has been derived from another SOP Instance.

Cardiac Synchronization

C.7.6.16.2.7

U

Frame Anatomy

C.7.6.16.2.8

M

Frame VOI LUT

C.7.6.16.2.10

M

Contrast/Bolus Usage

C.7.6.16.2.12

C - Required if the Enhanced Contrast/Bolus Module is present.

Pixel Intensity Relationship LUT

C.7.6.16.2.13

C - Required if Pixel Intensity Relationship (0028,1040) equals LOG. May be present otherwise.

Frame Pixel Shift

C.7.6.16.2.14

U

Patient Orientation in Frame

C.7.6.16.2.15

U

Frame Display Shutter

C.7.6.16.2.16

U

Respiratory Synchronization

C.7.6.16.2.17

U

Irradiation Event Identification

C.7.6.16.2.18

M

X-Ray Frame Characteristics

C.8.19.6.1

U

X-Ray Field of View

C.8.19.6.2

U

X-Ray Exposure Control Sensing Regions

C.8.19.6.3

U

X-Ray Frame Pixel Data Properties

C.8.19.6.4

M

X-Ray Frame Detector Parameters

C.8.19.6.5

C - Required if X-Ray Receptor Type (0018,9420) is present and equals DIGITAL_DETECTOR.

X-Ray Calibration Device Usage

C.8.19.6.6

U

X-Ray Object Thickness

C.8.19.6.7

U

X-Ray Frame Acquisition

C.8.19.6.8

U

X-Ray Positioner

C.8.19.6.10

U

X-Ray Table Position

C.8.19.6.11

U

X-Ray Collimator

C.8.19.6.12

U

X-Ray Geometry

C.8.19.6.14

U


A.48.4.1 Enhanced XRF Image Functional Group Macros Content Constraints

A.48.4.1.1 Frame Anatomy Functional Group Macro

For Anatomic Region Sequence (0008,2218) DCID 4042 “XA/XRF Anatomy Imaged” shall be used.

DICOM PS3.3 2024d - Information Object Definitions