DICOM PS3.3 2024d - Information Object Definitions |
---|
Table A.35.22-1 specifies the Modules of the Enhanced X-Ray Radiation Dose SR IOD.
Table A.35.22-1. Enhanced X-Ray Radiation Dose SR IOD Modules
The document shall be constructed from TID 10040 “Enhanced X-Ray Radiation Dose” invoked at the root node.
Value Type (0040,A040) in Content Sequence (0040,A730) of the SR Document Content Module is constrained to the following Enumerated Values (see Table C.17.3-7 for Value Type definitions):
Relationships between Content Items in the content of this IOD shall be conveyed by-value. Table A.35.22-2 specifies the relationship constraints of this IOD. See Table C.17.3-8 for Relationship Type definitions.
Table A.35.22-2. Relationship Content Constraints for Enhanced X-Ray Radiation Dose SR IOD
DICOM PS3.3 2024d - Information Object Definitions |
---|