DICOM PS3.3 2024d - Information Object Definitions |
---|
The Variable Modality LUT Softcopy Presentation State IOD specifies information that may be used to apply transformations to monochrome or pseudo-color images for which one or more Modality LUT transformations are intended for presentation control. This IOD supports presentation control for a set of images for which the Modality LUT varies on a frame-by-frame basis.
It includes capabilities from the Grayscale Softcopy Presentation State IOD and Pseudo-color Softcopy Presentation State IOD for specifying:
grayscale contrast transformations including Modality LUT and VOI LUT
a color palette to map the transformed grayscale values into pseudo-color
selection of the area of the image to display and whether to rotate or flip it
image and display relative annotations, including graphics, text and overlays
This IOD may only reference monochrome images or pseudo-color images, i.e. images with a Photometric Interpretation (0028,0004) of MONOCHROME1, MONOCHROME2 or PALETTE_COLOR. For other Photometric Interpretations, presentation behavior is not defined. See Section A.33.2 for the Color Softcopy Presentation State IOD, which allows for referencing color images, e.g. images with a Photometric Interpretation (0028,0004) of RGB.
DICOM PS3.3 2024d - Information Object Definitions |
---|