A.33.3.1 Pseudo-color Softcopy Presentation State IOD Description
The Pseudo-color Softcopy Presentation State IOD specifies information that may be used to present (display) images that are referenced from within the IOD.
It includes capabilities for specifying:
-
the output color space in PCS-Values
-
grayscale contrast transformations including modality and VOI LUT
-
a color palette to map the transformed grayscale values into pseudo-color
-
mask subtraction for multi-frame images
-
selection of the area of the image to display and whether to rotate or flip it
-
image and display relative annotations, including graphics, text and overlays
Note
This IOD does not support presentation control for a set of images for which the modality LUT varies on a frame-by-frame basis. Other Presentation State IODs (standard or private) may support such images.