DICOM PS3.3 2024d - Information Object Definitions |
---|
Table A.33.2-1 specifies the Modules of the Color Softcopy Presentation State IOD.
Table A.33.2-1. Color Softcopy Presentation State IOD Modules
C - Required if a Display Shutter is to be applied to referenced image(s) and the Bitmap Display Shutter Module is not present |
|||
C - Required if a Display Shutter is to be applied to referenced image(s) and the Display Shutter Module is not present |
|||
C - Required if Overlay is to be applied to referenced image(s) or the Bitmap Display Shutter Module is present. |
|||
C - Required if referenced image contains overlay data that is to be displayed or Presentation State Instance contains Overlay data other than Bitmap Shutter |
|||
C - Required if Graphic Annotations are to be applied to referenced image(s) |
|||
C - Required if rotation, flipping or magnification are to be applied to referenced image(s) |
|||
C - Required if Graphic Annotations or Overlays are to be applied to referenced image(s) |
|||
In the Color Softcopy Presentation State IOD, the Presentation State Shutter Module specializes some Attributes of the General Series Module, and the Presentation State Relationship Module specializes some Attributes of the Bitmap Display Shutter Module and Display Shutter Module.
DICOM PS3.3 2024d - Information Object Definitions |
---|