| DICOM PS3.22 2024c - Real-Time Communication |
|---|
The RTP Payload for audio and video Flows shall follow [SMPTE ST 2110-20] and [SMPTE ST 2110-30], respectively.
The RTP Payload for DICOM Metadata Flows (audio, video and rendition) shall follow [SMPTE ST 2110-10].
The RTP Payload for DICOM Metadata Flows consists of a DICOM dataset compliant with real-time communication.
The DICOM dataset is made of three parts:
the RTV Meta Information part. This part shall be present in each Grain.
the dynamic part containing information that varies over time (e.g., Origin Timestamp of the frame, Position of a probe, circle defining the eye. When it exists, this part shall be present in each Grain. The transmission rate of the dynamic part shall be identical to the rate of the associated Flow (e.g., one dataset per frame). This part is for the moment not applicable to DICOM Rendition Metadata.
the static part containing information that doesn't vary over time (e.g. Patient Name, Modality, …). This part will not be present in every Grain but shall be present at least in one Grain per second.
The receiver cannot process information received from a sender until it receives DICOM Metadata including the static part, so it has to be sent at least every second in order to avoid a longer wait by the receiver when "connected" to a sender.
The transmission rate of DICOM audio flows will be typically of the range of 48kHz. The transmission rate of DICOM video flows will be typically of the range of 60Hz. The transmission rate of the DICOM Rendition Metadata Flow shall be at least 1Hz. It may be appropriate to use a higher frequency if there is a need for tight synchronization of associated Flows from a device (e.g., two videos of a stereo pair).
| DICOM PS3.22 2024c - Real-Time Communication |
|---|