| DICOM PS3.18 2024c - Web Services |
|---|
The response shall have the following syntax:
version SP status-code SP reason-phrase CRLF
[retry-after CRLF]
CRLF
[payload]
Table 13.4.3-1 shows some common status codes corresponding to this transaction. See also Section 8.5 for additional status codes.
Table 13.4.3-1. Status Code Meaning
|
The origin server finished processing the storage commitment request; the payload describes in detail what referenced SOP Instances have been committed for safekeeping, and what Instances have not. |
||
|
The origin server has not finished processing the storage commitment request yet; there is no payload. The user agent is expected to follow-up with the Result Check transaction, as described in Section 13.5, to retrieve the result of the storage commitment request. |
||
|
The origin server cannot handle the storage commitment request because of errors in the request headers or parameters. |
||
|
The origin server cannot handle the storage commitment request because the provided transaction UID is already in use. |
||
|
The origin server cannot handle the storage commitment request; this may be a temporal or permanent state. |
| DICOM PS3.18 2024c - Web Services |
|---|