DICOM PS3.18 2024c - Web Services

J Storage Commitment Modules

J.1 Storage Commitment Request Module

Table J.1-1 specifies the Attributes of the Storage Commitment Request Module.

Table J.1-1.  Storage Commitment Request Module

Attribute Name

Tag

Type

Attribute Description

Referenced SOP Sequence

(0008,1199)

1C

The SOP Instances for which storage commitment is requested. One or more Items shall be included in this Sequence. Required if the Referenced Study Sequence (0008,1110) is absent.

>Include Table 10-11 “SOP Instance Reference Macro Attributes” in PS3.3

Referenced Study Sequence

(0008,1110)

1C

The Studies containing Instances for which storage commitment is requested. One or more Items shall be included in this Sequence. Required if the Referenced SOP Sequence (0008,1199) is absent.

>Study Instance UID

(0020,000D)

1

Uniquely identifies the referenced Study.

>Referenced Series Sequence

(0008,1115)

1

The Series containing Instances for which storage commitment is requested. One or more Items shall be included in this Sequence.

>>Series Instance UID

(0020,000E)

1

Uniquely identifies the referenced Series.

>>Referenced Instances by SOP Class Sequence

(0008,1112)

1

The SOP Instances for which storage commitment is requested grouped by SOP Class. One or more Items shall be included in this Sequence.

>>>Referenced SOP Class UID

(0008,1150)

1

Uniquely identifies the referenced SOP Class.

>>>Referenced Instance Sequence

(0008,114A)

1

The SOP Instances for which storage commitment is requested. One or more Items shall be included in this Sequence.

>>>>Referenced SOP Instance UID

(0008,1155)

1

Uniquely identifies the referenced SOP Instance.


DICOM PS3.18 2024c - Web Services