JJJJ.5.1.1 Example Multiple Physical Sources and Multiple Physical Detectors
This example shows an Effective Atomic Number image acquired on an acquisition device with multiple physical sources and multiple physical detectors, encoded as a CT Image IOD.
Table JJJJ.5.1.1-1. CT Image Module Attributes
|
Attribute Name
|
Tag
|
Values
|
|
Image Type
|
(0008,0008)
|
ORIGINAL\
PRIMARY\
AXIAL\
EFF_ATOMIC_NUM
|
|
Multi-energy CT Acquisition
|
(0018,9361)
|
YES
|
|
…
|
|
|
|
Rescale Intercept
|
(0028,1052)
|
-102.4
|
|
Rescale Slope
|
(0028,1053)
|
0.1
|
|
Rescale Type
|
(0028,1054)
|
Z_EFF
|
|
…
|
|
|
|
KVP
|
(0018,0060)
|
{null value because it is described below}
|
|
…
|
|
|
|
Distance Source to Detector
|
(0018,1110)
|
1000
|
|
Distance Source to Patient
|
(0018,1111)
|
500
|
|
Exposure Time
|
(0018,1150)
|
1000
|
|
Single Collimation Width
|
(0018,9306)
|
0.6
|
|
Total Collimation Width
|
(0018,9307)
|
38,4
|
|
…
|
|
|
|
Include ????
|
|
Table JJJJ.5.1.1-2. Multi-energy CT Image Attributes
Table JJJJ.5.1.1-3. Multi-energy CT X-Ray Source Macro Attributes
|
Attribute Name
|
Tag
|
Values
|
|
Multi-energy CT X-Ray Source Sequence
|
(0018,9365)
|
|
|
ITEM 1
|
|
>X-Ray Source Index
|
(0018,9366)
|
1
|
|
>X-Ray Source ID
|
(0018,9367)
|
Tube A
|
|
>Multi-energy Source Technique
|
(0018,9368)
|
CONSTANT_SOURCE
|
|
>Source Start DateTime
|
(0018,9369)
|
2018.05.01 13:22:03
|
|
>Source End DateTime
|
(0018,936A)
|
2018.05.01 13:22:20
|
|
>Generator Power
|
(0018,1170)
|
100
|
|
ITEM 2
|
|
>X-Ray Source Index
|
(0018,9366)
|
2
|
|
>X-Ray Source ID
|
(0018,9367)
|
Tube B
|
|
>Multi-energy Source Technique
|
(0018,9368)
|
CONSTANT_SOURCE
|
|
>Source Start DateTime
|
(0018,9369)
|
2018.05.01 13:22:03
|
|
>Source End DateTime
|
(0018,936A)
|
2018.05.01 13:22:20
|
|
>Generator Power
|
(0018,1170)
|
100
|
Table JJJJ.5.1.1-4. Multi-energy CT X-Ray Detector Macro Attributes
|
Attribute Name
|
Tag
|
Values
|
|
Multi-energy CT X-Ray Detector Sequence
|
(0018,936F)
|
|
|
ITEM 1
|
|
>X-Ray Detector Index
|
(0018,9370)
|
1
|
|
>X-Ray Detector ID
|
(0018,9371)
|
Detector A
|
|
>Multi-energy Detector Type
|
(0018,9372)
|
INTEGRATING
|
|
>X-Ray Detector Label
|
(0018,9373)
|
High-Energy
|
|
>Nominal Max Energy
|
(0018,9374)
|
150
|
|
>Nominal Min Energy
|
(0018,9375)
|
35
|
|
>Effective Bin Energy
|
(0018,936E)
|
90
|
|
ITEM 2
|
|
>X-Ray Detector Index
|
(0018,9370)
|
2
|
|
>X-Ray Detector ID
|
(0018,9371)
|
Detector B
|
|
>Multi-energy Detector Type
|
(0018,9372)
|
INTEGRATING
|
|
>X-Ray Detector Label
|
(0018,9373)
|
Low-Energy
|
|
>Nominal Max Energy
|
(0018,9374)
|
100
|
|
>Nominal Min Energy
|
(0018,9375)
|
35
|
|
>Effective Bin Energy
|
(0018,936E)
|
60
|
Table JJJJ.5.1.1-5. Multi-energy CT Path Macro Attributes
|
Attribute Name
|
Tag
|
Values
|
|
Multi-energy CT Path Sequence
|
(0018,9379)
|
|
|
ITEM 1
|
|
>Multi-energy CT Path Index
|
(0018,937A)
|
1
|
|
>X-Ray Source Index
|
(0018,9366)
|
1
|
|
>X-Ray Detector Index
|
(0018,9370)
|
1
|
|
ITEM 2
|
|
>Multi-energy CT Path Index
|
(0018,937A)
|
2
|
|
>X-Ray Source Index
|
(0018,9366)
|
2
|
|
>X-Ray Detector Index
|
(0018,9370)
|
2
|
Table JJJJ.5.1.1-6. CT Exposure Macro Attributes
|
Attribute Name
|
Tag
|
Values
|
|
CT Exposure Sequence
|
(0018,9321)
|
|
|
ITEM 1
|
|
>Referenced X-Ray Source Index
|
(0018,9377)
|
1
|
|
>Exposure Time in ms
|
(0018,9328)
|
1000
|
|
>X-Ray Tube Current in mA
|
(0018,9330)
|
500
|
|
>Exposure in mAs
|
(0018,9332)
|
500
|
|
>Exposure Modulation Type
|
(0018,9323)
|
CD4D
|
|
>CTDIvol
|
(0018,9345)
|
5
|
|
ITEM 2
|
|
>Referenced X-Ray Source Index
|
(0018,9377)
|
2
|
|
>Exposure Time in ms
|
(0018,9328)
|
1000
|
|
>X-Ray Tube Current in mA
|
(0018,9330)
|
250
|
|
>Exposure in mAs
|
(0018,9332)
|
250
|
|
>Exposure Modulation Type
|
(0018,9323)
|
CD4D
|
|
>CTDIvol
|
(0018,9345)
|
5
|
Table JJJJ.5.1.1-7. CT X-Ray Details Sequence Macro Attributes
|
Attribute Name
|
Tag
|
Values
|
|
CT X-Ray Details Sequence
|
(0018,9325)
|
|
|
ITEM 1
|
|
>Referenced Path Index
|
(0018,9378)
|
1
|
|
>KVP
|
(0018,0060)
|
150
|
|
>Focal Spot(s)
|
(0018,1190)
|
1.2
|
|
>Filter Type
|
(0018,1160)
|
WEDGE2
|
|
>Filter Material
|
(0018,7050)
|
MIXED
|
|
ITEM 2
|
|
>Referenced Path Index
|
(0018,9378)
|
2
|
|
>KVP
|
(0018,0060)
|
100
|
|
>Focal Spot(s)
|
(0018,1190)
|
1.2
|
|
>Filter Type
|
(0018,1160)
|
WEDGE2+FLAT
|
|
>Filter Material
|
(0018,7050)
|
TIN
|
Table JJJJ.5.1.1-8. CT Acquisition Details Macro Attributes
|
Attribute Name
|
Tag
|
Values
|
|
CT Acquisition Details Sequence
|
(0018,9304)
|
|
|
ITEM 1
|
|
>Referenced Path Index
|
(0018,9378)
|
1
|
|
>Rotation Direction
|
(0018,1140)
|
CW
|
|
>Revolution Time
|
(0018,9305)
|
0.5
|
|
>Single Collimation Width
|
(0018,9306)
|
0.6
|
|
>Total Collimation Width
|
(0018,9307)
|
38.4
|
|
>Table Height
|
(0018,1130)
|
88.5
|
|
>Gantry/Detector Tilt
|
(0018,1120)
|
0
|
|
>Data Collection Diameter
|
(0018,0090)
|
500
|
|
ITEM 2
|
|
>Referenced Path Index
|
(0018,9378)
|
2
|
|
>Rotation Direction
|
(0018,1140)
|
CW
|
|
>Revolution Time
|
(0018,9305)
|
0.5
|
|
>Single Collimation Width
|
(0018,9306)
|
0.6
|
|
>Total Collimation Width
|
(0018,9307)
|
38.4
|
|
>Table Height
|
(0018,1130)
|
88.5
|
|
>Gantry/Detector Tilt
|
(0018,1120)
|
0
|
|
>Data Collection Diameter
|
(0018,0090)
|
350
|
Table JJJJ.5.1.1-9. CT Geometry Macro Attributes
|
Attribute Name
|
Tag
|
Values
|
|
CT Geometry Sequence
|
(0018,9312)
|
|
|
ITEM 1
|
|
>Referenced Path Index
|
(0018,9378)
|
1\2
|
|
>Distance Source to Detector
|
(0018,1110)
|
1000
|
|
>Distance Source to Data Collection Center
|
(0018,9335)
|
500
|
Table JJJJ.5.1.1-10. Multi-energy CT Processing Attributes
|
Attribute Name
|
Tag
|
Values
|
|
Decomposition Method
|
(0018,937E)
|
HYBRID
|
|
Decomposition Description
|
(0018,937F)
|
iBHC + MAT DECOMP
|
JJJJ.5.1.2 Example Single Source Multi-layer Detector
This example shows a type Effective Atomic Number image acquired on an acquisition device with a single source and multi-layer detector.
Table JJJJ.5.1.2-1. CT Image Module Attributes
|
Attribute Name
|
Tag
|
Values
|
|
Image Type
|
(0008,0008)
|
ORIGINAL\PRIMARY\AXIAL\EFF_ATOMIC_NUM
|
|
Multi-energy CT Acquisition
|
(0018,9361)
|
YES
|
|
…
|
|
|
|
Rescale Intercept
|
(0028,1052)
|
0
|
|
Rescale Slope
|
(0028,1053)
|
1.3
|
|
Rescale Type
|
(0028,1054)
|
10^-2 Z_EFF
|
|
…
|
|
|
|
KVP
|
(0018,0060)
|
{null value because it is described below}
|
|
…
|
|
|
|
Scan Options
|
(0018,0022)
|
AXIAL
|
|
Data Collection Diameter
|
(0018,0090)
|
500
|
|
…
|
|
|
|
Distance Source to Detector
|
(0018,1110)
|
1040
|
|
Distance Source to Patient
|
(0018,1111)
|
570
|
|
Exposure Time
|
(0018,1150)
|
750
|
|
X-Ray Tube Current
|
(0018,1151)
|
440
|
|
Exposure
|
(0018,1152)
|
330
|
|
Single Collimation Width
|
(0018,9306)
|
0.625
|
|
Total Collimation Width
|
(0018,9307)
|
20.0
|
|
…
|
|
|
|
Include ????
|
|
Table JJJJ.5.1.2-2. Multi-energy CT Image Attributes
Table JJJJ.5.1.2-3. Multi-energy CT X-Ray Source Macro Attributes
|
Attribute Name
|
Tag
|
Values
|
|
Multi-energy CT X-Ray Source Sequence
|
(0018,9365)
|
|
|
ITEM 1
|
|
>X-Ray Source Index
|
(0018,9366)
|
1
|
|
>X-Ray Source ID
|
(0018,9367)
|
Tube A
|
|
>Multi-energy Source Technique
|
(0018,9368)
|
CONSTANT_SOURCE
|
|
>Source Start DateTime
|
(0018,9369)
|
2018.05.01 13:22:03
|
|
>Source End DateTime
|
(0018,936A)
|
2018.05.01 13:22:20
|
Table JJJJ.5.1.2-4. Multi-energy CT X-Ray Detector Macro Attributes
|
Attribute Name
|
Tag
|
Values
|
|
Multi-energy CT X-Ray Detector Sequence
|
(0018,936F)
|
|
|
ITEM 1
|
|
>X-Ray Detector Index
|
(0018,9370)
|
1
|
|
>X-Ray Detector ID
|
(0018,9371)
|
Detector A
|
|
>Multi-energy Detector Type
|
(0018,9372)
|
MULTILAYER
|
|
>X-Ray Detector Label
|
(0018,9373)
|
High-Energy
|
|
ITEM 2
|
|
>X-Ray Detector Index
|
(0018,9370)
|
2
|
|
>X-Ray Detector ID
|
(0018,9371)
|
Detector A
|
|
>Multi-energy Detector Type
|
(0018,9372)
|
MULTILAYER
|
|
>X-Ray Detector Label
|
(0018,9373)
|
Low-Energy
|
Table JJJJ.5.1.2-5. Multi-energy CT Path Macro Attributes
|
Attribute Name
|
Tag
|
Values
|
|
Multi-energy CT Path Sequence
|
(0018,9379)
|
|
|
ITEM 1
|
|
>Multi-energy CT Path Index
|
(0018,937A)
|
1
|
|
>X-Ray Source Index
|
(0018,9366)
|
1
|
|
>X-Ray Detector Index
|
(0018,9370)
|
1
|
|
ITEM 2
|
|
>Multi-energy CT Path Index
|
(0018,937A)
|
2
|
|
>X-Ray Source Index
|
(0018,9366)
|
1
|
|
>X-Ray Detector Index
|
(0018,9370)
|
2
|
Table JJJJ.5.1.2-6. CT Exposure Macro Attributes
|
Attribute Name
|
Tag
|
Values
|
|
CT Exposure Sequence
|
(0018,9321)
|
|
|
ITEM 1
|
|
>Referenced X-Ray Source Index
|
(0018,9377)
|
1
|
|
>Exposure Time in ms
|
(0018,9328)
|
750
|
|
>X-Ray Tube Current in mA
|
(0018,9330)
|
440
|
|
>Exposure in mAs
|
(0018,9332)
|
330
|
|
>Exposure Modulation Type
|
(0018,9323)
|
NONE
|
|
>CTDIvol
|
(0018,9345)
|
34.9
|
Table JJJJ.5.1.2-7. CT X-Ray Details Sequence Macro Attributes
|
Attribute Name
|
Tag
|
Values
|
|
CT X-Ray Details Sequence
|
(0018,9325)
|
|
|
ITEM 1
|
|
>Referenced Path Index
|
(0018,9378)
|
1\2
|
|
>KVP
|
(0018,0060)
|
120
|
|
>Focal Spot(s)
|
(0018,1190)
|
1.4
|
|
>Filter Type
|
(0018,1160)
|
NONE
|
|
>Filter Material
|
(0018,7050)
|
|
Table JJJJ.5.1.2-8. CT Acquisition Details Macro Attributes
|
Attribute Name
|
Tag
|
Values
|
|
CT Acquisition Details Sequence
|
(0018,9304)
|
|
|
ITEM 1
|
|
>Referenced Path Index
|
(0018,9378)
|
1
|
|
>Rotation Direction
|
(0018,1140)
|
CW
|
|
>Revolution Time
|
(0018,9305)
|
0.75
|
|
>Single Collimation Width
|
(0018,9306)
|
0.625
|
|
>Total Collimation Width
|
(0018,9307)
|
20.0
|
|
>Table Height
|
(0018,1130)
|
88.5
|
|
>Gantry/Detector Tilt
|
(0018,1120)
|
0
|
|
>Data Collection Diameter
|
(0018,0090)
|
500
|
Table JJJJ.5.1.2-9. CT Geometry Macro Attributes
|
Attribute Name
|
Tag
|
Values
|
|
CT Geometry Sequence
|
(0018,9312)
|
|
|
ITEM 1
|
|
>Referenced Path Index
|
(0018,9378)
|
1
|
|
>Distance Source to Detector
|
(0018,1110)
|
1140
|
|
>Distance Source to Data Collection Center
|
(0018,9335)
|
570
|
Table JJJJ.5.1.2-10. Multi-energy CT Processing Attributes
|
Attribute Name
|
Tag
|
Values
|
|
Decomposition Method
|
(0018,937E)
|
PROJECTION_BASED
|
|
Decomposition Description
|
(0018,937F)
|
Photo-Electric / Compton Scattering Decomposition
|