DICOM PS3.17 2024c - Explanatory Information

CCCC.5 Example

Nest

Relationship

Value Type

Concept Name

Example Value

CONTAINER

Adult Echocardiography Procedure Report

>

CONTAINS

CONTAINER

(125301, DCM, "Pre-coordinated Measurements")

>>

CONTAINS

NUM

(79969-2, LN, "Interventricular septum diastolic dimension")

1.00 (cm,UCUM,"cm")

>>>

HAS PROPERTIES

TEXT

(125309, DCM, "Short Label")

"IVSd (2D) "

>>

CONTAINS

NUM

(79991-6, LN, "Left ventricular ejection fraction biplane (MOD) ")

70.3 (%,UCUM,"%")

>>>

HAS PROPERTIES

TEXT

(125309, DCM, "Short Label")

"LV EF (MOD) "

>>

CONTAINS

NUM

(79996-5, LN, "Left ventricular end diastolic volume biplane (MOD) ")

118 (ml,UCUM,"ml")

>>>

HAS PROPERTIES

TEXT

(125309, DCM, "Short Label")

"LV EDV (MOD) "

>>

CONTAINS

NUM

(80001-1, LN, "Left ventricular end systolic volume biplane (MOD) ")

35.0 (ml,UCUM,"ml")

>>>

HAS PROPERTIES

TEXT

(125309, DCM, "Short Label")

"LV ESV (MOD) "

>>

CONTAINS

NUM

(80007-8, LN, "Left ventricular internal diastolic dimension - 2D")

5.00 (cm,UCUM,"cm")

>>>

HAS PROPERTIES

CODE

(121404, DCM, "Selection Status")

(121410, DCM, "User chosen value")

>>>

HAS PROPERTIES

TEXT

(125309, DCM, "Short Label")

"LVIDd (2D) "

>>

CONTAINS

NUM

(80007-8, LN, "Left ventricular internal diastolic dimension - 2D")

5.50 (cm,UCUM,"cm")

>>>

HAS PROPERTIES

TEXT

(125309, DCM, "Short Label")

"LVIDd (2D) "

>>

CONTAINS

NUM

(80007-8, LN, "Left ventricular internal diastolic dimension - 2D")

6.00 (cm,UCUM,"cm")

>>>

HAS PROPERTIES

TEXT

(125309, DCM, "Short Label")

"LVIDd (2D) "

>>

CONTAINS

NUM

(80011-0, LN, "Left ventricular internal systolic dimension - 2D")

3.00 (cm,UCUM,"cm")

>>>

HAS PROPERTIES

TEXT

(125309, DCM, "Short Label")

"LVIDs (2D) "

>>

CONTAINS

NUM

(80031-8, LN, "Left ventricular posterior wall diastolic thickness")

1.00 (cm,UCUM,"cm")

>>>

HAS PROPERTIES

TEXT

(125309, DCM, "Short Label")

"LVPWd (2D) "

>>

CONTAINS

NUM

(80068-0, LN, "Mitral valve area (Planimetry) ")

4.82 (cm2,UCUM,"cm2")

>>>

HAS PROPERTIES

TEXT

(125309, DCM, "Short Label")

"MV Area (Planim) "

>

CONTAINS

CONTAINER

(125302, DCM, "Post-coordinated Measurements")

>>

CONTAINS

NUM

(LVSIMOD,99CompanyName,"Left Ventricle Stroke Index (MOD) ")

39 (ml/m2,UCUM,"ml/m2")

>>>

HAS CONCEPT MOD

CODE

(125306, DCM, "Measurement Type")

(125313, DCM, "Indexed")

>>>

HAS CONCEPT MOD

CODE

(363698007, SCT, "Finding Site")

(87878005, SCT, "Left Ventricle")

>>>

HAS CONCEPT MOD

CODE

(125305, DCM, "Finding Observation Type")

(44324008, SCT, "Hemodynamic Measurements")

>>>

HAS CONCEPT MOD

CODE

(125307, DCM, "Measured Property")

(90096001, SCT, "Stroke Volume")

>>>

HAS CONCEPT MOD

CODE

(370129005, SCT, "Measurement Method")

(125207, DCM, "Method of Disks Biplane")

>>>

HAS CONCEPT MOD

CODE

(399264008, SCT, "Image Mode")

(399064001, SCT, "2D Mode")

>>>

HAS CONCEPT MOD

CODE

(125308, DCM, "Measurement Divisor")

(8277-6, LN, "Body Surface Area")

>>>

HAS PROPERTIES

TEXT

(125309, DCM, "Short Label")

"LV SI (MOD) "

>>

CONTAINS

NUM

(29469-4, LN, "Left Atrium Antero-posterior Systolic Dimension")

3.0 (cm,UCUM,"cm")

>>>

HAS CONCEPT MOD

CODE

(125306, DCM, "Measurement Type")

(125316, DCM, "Directly measured")

>>>

HAS CONCEPT MOD

CODE

(363698007, SCT, "Finding Site")

(82471001, SCT, "Left Atrium")

>>>

HAS CONCEPT MOD

CODE

(125305, DCM, "Finding Observation Type")

(125311, DCM, "Structure of the Finding Site")

>>>

HAS CONCEPT MOD

CODE

(125307, DCM, "Measured Property")

(81827009, SCT, "Diameter")

>>>

HAS CONCEPT MOD

CODE

(370129005, SCT, "Measurement Method")

(122675, DCM, "Anterior-Posterior")

>>>

HAS CONCEPT MOD

CODE

(399264008, SCT, "Image Mode")

(399064001, SCT, "2D Mode")

>>>

HAS CONCEPT MOD

CODE

(272518008, SCT, "Cardiac Cycle Point")

(416430001, SCT, "End Systole")

>>>

HAS PROPERTIES

TEXT

(125309, DCM, "Short Label")

"LA Dimen (2D) "

>

CONTAINS

CONTAINER

(125303, DCM, "Adhoc Measurements")

>>

CONTAINS

NUM

(385673002, SCT, "Interval")

15.0 (ms,UCUM,"ms")

>>>

HAS PROPERTIES

TEXT

(125309, DCM, "Short Label")

"MV Jet Duration"

>>

CONTAINS

NUM

(1483009, SCT, "Angle")

27.0 (deg,UCUM,"deg")

>>>

HAS PROPERTIES

TEXT

(125309, DCM, "Short Label")

"MV Leaf Angle"

DICOM PS3.17 2024c - Explanatory Information