DICOM PS3.17 2024c - Explanatory Information

AAAA.3.2 Acme

The first part of this example is shown above in Table AAAA.3-2.

The anatomical extent is defined in the reconstruction to represent the dataset of interest to the clinical trial. The extent was not defined in the localizer or acquisition. Sites are welcome to reflect their local practice in the localizer and acquisition extent as long as they permit production of the reconstruction as specified here.

Table AAAA.3-2. CT Tumor Volumetric Measurement - Details - Acme

Attribute

Tag

Value

Model Specification Sequence

(0018,9912)

>Manufacturer

(0008,0070)

ACME

>Manufacturer's Related Model Group

(0008,0222)

Ultimate

>Software Versions

(0018,1020)

V3.1

Acquisition Protocol Element Specification Sequence

(0018,991F)

>Protocol Element Number

(0018,9921)

1

>Parameters Specification Sequence

(0018,9913)

>>See Table AAAA.3-2a First Acquisition Protocol Element Specification

>Protocol Element Number

(0018,9921)

2

>Parameters Specification Sequence

(0018,9913)

>>See Table AAAA.3-2b “Second Acquisition Protocol Element Specification”

Reconstruction Protocol Element Specification Sequence

(0018,9933)

>Protocol Element Number

(0018,9921)

1

>Parameters Specification Sequence

(0018,9913)

>>See Table AAAA.3-2c “First Reconstruction Protocol Element Specification”


The following tables reflect the semantic contents of constraint sequences but not the actual structure of the IOD. The centered rows in italics clarify the context of the constrained Attributes that follow by indicating which sequence in the performed module contains the constrained Attribute (as specified in the Selector Sequence Pointer).

Table AAAA.3-2a. First Acquisition Protocol Element Specification

Attribute

Selector Attribute

Selector Value Number

Selector Sequence Pointer

Selector Sequence Pointer Items

Constraint Type

Constraint Value

Acquisition Protocol Element Sequence (0018,9920)

Protocol Element Name

(0018,9922)

1

(0018,9920)

1

EQUAL

Localizer: Lateral

Acquisition Type

(0018,9302)

1

(0018,9920)

1

EQUAL

CONSTANT_ANGLE

Tube Angle

(0018,9303)

1

(0018,9920)

1

EQUAL

90

Acquisition Motion

(0018,9930)

1

(0018,9920)

1

EQUAL

FORWARD

>CT X-Ray Details Sequence (0018,9325) - First Beam

Beam Number

(300A,00C0)

1

(0018,9920), (0018,9325)

1\1

EQUAL

1

KVP

(0018,0060)

1

(0018,9920), (0018,9325)

1\1

EQUAL

120

X-ray Tube Current in mA

(0018,9330)

1

(0018,9920), (0018,9325)

1\1

EQUAL

50


Table AAAA.3-2b. Second Acquisition Protocol Element Specification

Attribute

Selector Attribute

Selector Value Number

Selector Sequence Pointer

Selector Sequence Pointer Items

Constraint Type

Constraint Value

Acquisition Protocol Element Sequence (0018,9920)

Protocol Element Name

(0018,9922)

1

(0018,9920)

2

EQUAL

Helical

Acquisition Type

(0018,9302)

1

(0018,9920)

2

EQUAL

SPIRAL

Revolution Time

(0018,9305)

1

(0018,9920)

2

EQUAL

0.5

Single Collimation Width

(0018,9306)

1

(0018,9920)

2

EQUAL

0.75

Total Collimation Width

(0018,9307)

1

(0018,9920)

2

EQUAL

48

Table Speed

(0018,9309)

1

(0018,9920)

2

EQUAL

27

>CT X-Ray Details Sequence (0018,9325) - First Beam

Beam Number

(300A,00C0)

1

(0018,9920), (0018,9325)

2\1

EQUAL

1

KVP

(0018,0060)

1

(0018,9920), (0018,9325)

2\1

EQUAL

120

Exposure in mAs

(0018,9332)

1

(0018,9920), (0018,9325)

2\1

RANGE_INCL

100, 260

Respiratory Motion Compensation Technique

(0018,9170)

1

(0018,9920), (0018,9325)

2\1

EQUAL

BREATH_HOLD


Table AAAA.3-2c. First Reconstruction Protocol Element Specification

Attribute

Selector Attribute

Selector Value Number

Selector Sequence Pointer

Selector Sequence Pointer Items

Constraint Type

Constraint Value

Reconstruction Protocol Element Sequence (yym8, m9x1)

Protocol Element Name

(0018,9922)

1

(0018,9934)

1

EQUAL

"Transverse"

Source Acquisition Protocol Element Number

(0018,9938)

1

(0018,9934)

1

EQUAL

2

Source Acquisition Beam Number

(0018,9939)

1

(0018,9934)

1

EQUAL

1

Reconstruction Algorithm

(0018,9315)

1

(0018,9934)

1

EQUAL

FILTER_BACK_PROJ

Convolution Kernel

(0018,1210)

1

(0018,9934)

1

EQUAL

"B1"

Convolution Kernel Group

(0018,9316)

1

(0018,9934)

1

EQUAL

"LUNG"

Reconstruction Pixel Spacing

(0018,9322)

1

(0018,9934)

1

RANGE_INCL

0.55, 0.75

Slice Thickness

(0018,0050)

1

(0018,9934)

1

EQUAL

1.0

Spacing Between Slices

(0018,0088)

1

(0018,9934)

1

EQUAL

1.0

>Reconstruction Start Location Sequence (0018,993B)

Reference Location Label

(0018,9900)

1

(0018,9934), (0018,993B)

1\1

EQUAL

"Top of Shoulders"

Reference Basis Code Sequence

(0018,9902)

1

(0018,9934), (0018,993B)

1\1

EQUAL

(16982005, SCT, "Shoulder region structure")

Reference Geometry Code Sequence

(0018,9903)

1

(0018,9934), (0018,993B)

1\1

EQUAL

(128120, DCM, "Plane through Superior Extent")

>Reconstruction End Location Sequence (0018,993C)

Reference Location Label

(0018,9900)

1

(0018,9934), (0018,993C)

1\1

EQUAL

"Mid-liver"

Reference Basis Code Sequence

(0018,9902)

1

(0018,9934), (0018,993C)

1\1

EQUAL

(10200004, SCT, "Liver")

Reference Geometry Code Sequence

(0018,9903)

1

(0018,9934), (0018,993C)

1\1

EQUAL

(128130, DCM, "Plane through Center")


DICOM PS3.17 2024c - Explanatory Information