DICOM PS3.16 2024c - Content Mapping Resource

Ultrasound Shear Wave Elastography Templates 

TID 5401 Ultrasound Shear Wave Elastography Section

This section template incorporates a set of measurements for assessing tissue elasticity characteristics based on shear wave speed.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 5401. Ultrasound Shear Wave Elastography Section

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

DT (59776-5, LN, "Findings")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (121058, DCM, "Procedure Reported")

1

M

DT (448764002, SCT, "Ultrasound elastography (procedure)")

3

>

HAS CONCEPT MOD

CODE

EV (363698007, SCT, "Finding Site")

1

M

BCID 12321 “Elastography Site”

4

>>

HAS CONCEPT MOD

CODE

EV (272741003, SCT, "Laterality")

1

U

DCID 244 “Laterality”

5

>

HAS ACQ CONTEXT

CODE

EV (399264008, SCT, "Image Mode")

1

U

BCID 12224 “Ultrasound Image Mode”

6

>

HAS ACQ CONTEXT

CODE

EV (111031, DCM, "Image View")

1

U

BCID 5 “Transducer Approach”

7

>>

HAS ACQ CONTEXT

CODE

EV (111032, DCM, "Image View Modifier")

1-n

U

BCID 6 “Transducer Orientation”

BCID 7 “Ultrasound Beam Path”

8

>

HAS CONCEPT MOD

CODE

EV (130759, DCM, "Shear Wave Detection Method")

1

U

DCID 12324 “Shear Wave Detection Method”

9

>

CONTAINS

CONTAINER

DT (55112-7, LN, "Summary")

1

M

10

>>

CONTAINS

NUM

EV (130611, DCM, "Shear Wave Speed")

1

M

UNITS = EV (m/s, UCUM, "m/s")

11

>>>

HAS PROPERTIES

NUM

EV (386136009, SCT, "Standard deviation")

1

U

UNITS = EV (m/s, UCUM, "m/s")

12

>>>

HAS PROPERTIES

NUM

EV (373099004, SCT, "Median")

1

U

UNITS = EV (m/s, UCUM, "m/s")

13

>>>

HAS PROPERTIES

NUM

EV (130614, DCM, "Interquartile Range of population")

1

U

UNITS = EV (m/s, UCUM, "m/s")

14

>>>

HAS PROPERTIES

NUM

EV (130615, DCM, "Interquartile Range to Median Ratio of population")

1

M

UNITS = EV ({ratio}, UCUM, "ratio")

15

>>

CONTAINS

NUM

EV (110830, DCM, "Elasticity")

1

M

UNITS = EV (kPa, UCUM, "kPa")

16

>>>

HAS PROPERTIES

NUM

EV (386136009, SCT, "Standard deviation")

1

U

UNITS = EV (kPa, UCUM, "kPa")

17

>>>

HAS PROPERTIES

NUM

EV (373099004, SCT, "Median")

1

U

UNITS = EV (kPa, UCUM, "kPa")

18

>>>

HAS PROPERTIES

NUM

EV (130614, DCM, "Interquartile Range of population")

1

U

UNITS = EV (kPa, UCUM, "kPa")

19

>>>

HAS PROPERTIES

NUM

EV (130615, DCM, "Interquartile Range to Median Ratio of population")

1

M

UNITS = EV ({ratio}, UCUM, "ratio")

20

>>

CONTAINS

NUM

EV (130612, DCM, "Shear Wave Dispersion Slope")

1

U

UNITS = EV (m/s/kHz, UCUM, "m/s/kHz")

21

>>>

HAS PROPERTIES

NUM

EV (386136009, SCT, "Standard deviation")

1

U

UNITS = EV (m/s/kHz, UCUM, "m/s/kHz")

22

>>>

HAS PROPERTIES

NUM

EV (373099004, SCT, "Median")

1

U

UNITS = EV (m/s/kHz, UCUM, "m/s/kHz")

23

>>>

HAS PROPERTIES

NUM

EV (130614, DCM, "Interquartile Range of population")

1

U

UNITS = EV (m/s/kHz, UCUM, "m/s/kHz")

24

>>>

HAS PROPERTIES

NUM

EV (130615, DCM, "Interquartile Range to Median Ratio of population")

1

M

UNITS = EV ({ratio}, UCUM, "ratio")

25

>

CONTAINS

CONTAINER

DT (125007, DCM, "Measurement Group")

1-n

M

26

>>

HAS OBS CONTEXT

TEXT

EV (125010, DCM, "Identifier")

1

M

27

>>

HAS CONCEPT MOD

CODE

EV (363698007, SCT, "Finding Site")

1

U

BCID 12322 “Elastography Measurement Site”

28

>>

INCLUDE

DTID 5402 “Shear Wave Elastography Measurement”

1

M

29

>

CONTAINS

CONTAINER

DT (130755, DCM, "Reference Measurement Group")

1

U

30

>>

HAS CONCEPT MOD

CODE

EV (363698007, SCT, "Finding Site")

1

U

BCID 12322 “Elastography Measurement Site”

31

>>

INCLUDE

DTID 5402 “Shear Wave Elastography Measurement”

1

M


Content Item Descriptions

Row 10

The nominal shear wave speed for the tissue of interest. How this nominal value is derived from the measurements in Row 25 is implementation dependent.

Rows 11-14

These values are a statistical characterization of shear wave speed measurements in Row 25 from which the value in Row 10 was derived.

Row 15

The nominal Elasticity for the tissue of interest. How this nominal value is derived from the measurements in Row 25 is implementation dependent.

Rows 16-19

These values are a statistical characterization of Elasticity measurements in Row 25 from which the value in Row 15 was derived.

Row 20

The nominal Shear Wave Dispersion Slope for the tissue of interest. How this nominal value is derived from the measurements in Row 25 is implementation dependent.

Rows 21-24

These values are a statistical characterization of Shear Wave Dispersion Slope measurements in Row 25 from which the value in Row 20 was derived.

Row 26

The identifier is used to distinguish the different measurement ROIs. For example, the identifier might be used as a column header when displaying the measurements from all the ROIs in a table.

Row 27

The specific location of a measurement, e.g. the specific segment of the liver, may be recorded here. If this row is absent, the measurement is presumed to be somewhere within the organ/tissue identified in Row 3.

Row 29

In Shear Wave Elastography, the Reference Measurement Group is used to record measured values for a reference region against which the measurements in Row 25 may be compared.

Row 30

A coded description of the location of the reference measurement. If this row is absent, the reference measurement is presumed to be some conventional location appropriate for the organ/tissue identified in Row 3. If the SCOORD in TID 5402 is present, that indicates the spatial location.

DICOM PS3.16 2024c - Content Mapping Resource