| DICOM PS3.16 2024c - Content Mapping Resource |
|---|
This Template collects a composite feature for a lesion, non-lesion object, or correlation of related objects (see TID 4121 “Colon CAD Findings Summary”). The details of the composition are expressed in the Colon CAD Composite Feature Body (see TID 4126 “Colon CAD Composite Feature Body”). The data from which the details are inferred, are expressed in the Composite Features (see TID 4125 “Colon CAD Composite Feature”) and/or Single Image Findings (see TID 4127 “Colon CAD Single Image Finding”), of which there may be several.
Table TID 4125. Colon CAD Composite Feature
|
IFF value of Row 3 = (111151, DCM, "Presentation Optional") and Row 1 of TID 4023 “CAD Operating Points” is present for the feature identified in Row 1. |
UNITS = DT ({1:n}, UCUM, "range: 1:n"), where n is the maximum specified in Row 1 of TID 4023 “CAD Operating Points” for the feature identified in Row 1. Value is restricted to being an integer. |
|||||||
|
Shall be present IFF this feature is duplicated from a different report than its parent. |
||||||||
|
This Content Item constrains the SCP receiving the Colon CAD SR IOD in its use of the contents of this Template and its Target Content Items. Colon CAD devices may opt to use data marked "Not for Presentation" or "Presentation Optional" as input to subsequent colon CAD processing steps. Refer to Section O.2 “Structured Reporting Storage SOP Class SCU and SCP Behavior” in PS3.4 . |
| DICOM PS3.16 2024c - Content Mapping Resource |
|---|