| DICOM PS3.16 2024c - Content Mapping Resource |
|---|
The Atrial Pressure Measurement Template consists of a CONTAINER, with an acquisition context of the measurement anatomic site, and containing a-wave, v-wave, and mean measurements. This implies that the name of the measurement is effectively post-coordinated from the measurement name, the Hemodynamic Measurement Group container (procedure phase) name, and the acquisition context (finding site) value.
Table TID 3505. Atrial Pressure Measurement
|
$LocationName = EV (363698007, SCT, "Finding Site") $LocationValue = DCID 3608 “Atrial Source Location” |
||||||||
|
$Measurement = EV (109016, DCM, "A-wave peak pressure") $Units = DCID 3500 “Pressure Unit” |
||||||||
|
$Measurement = EV (109034, DCM, "V-wave peak pressure") $Units = DCID 3500 “Pressure Unit” |
||||||||
|
$Measurement = EV (6797001, SCT, "Mean blood pressure") $Units = DCID 3500 “Pressure Unit” |
||||||||
| DICOM PS3.16 2024c - Content Mapping Resource |
|---|