| DICOM PS3.16 2024c - Content Mapping Resource |
|---|
Table TID 3309. Stress Echo Measurement Group
|
$Measurement = DCID 12200 “Echocardiography Left Ventricle Measurement” |
||||||||
|
IF Row 3 measurement concept is in CID 12222 “Orifice Flow Property” |
||||||||
|
$Measurement = DCID 12211 “Echocardiography Aortic Valve Measurement” |
||||||||
|
IF Row 5 measurement concept is in CID 12222 “Orifice Flow Property” |
||||||||
|
$Measurement = DCID 12207 “Echocardiography Mitral Valve Measurement” |
||||||||
|
IF Row 7 measurement concept is in CID 12222 “Orifice Flow Property” |
||||||||
|
$Measurement = DCID 12208 “Echocardiography Tricuspid Valve Measurement” |
||||||||
|
IF Row 9 measurement concept is in CID 12222 “Orifice Flow Property” |
||||||||
|
$Procedure = DT (35757004, SCT, "Echocardiography for Determining Ventricular Contraction") |
|
These invocations of TID 5203 “Echo Measurement” do not include an inherited Findings Site concept, for example as in the invocations of TID 5203 “Echo Measurement” from TID 5202 “Echo Section”. Echo measurements that do not have the associated Finding Site pre-coordinated in the measurement concept (i.e., the orifice flow measurements of CID 12222 “Orifice Flow Property”), shall have the Finding Site explicitly post-coordinated with a Concept Modifier (Rows 4, 6, 8, and 10). This Template does not include the concept of an Image Library, for example as used in TID 5200 “Echocardiography Procedure Report”. Image Content Items in the Echo Measurement Template shall be included with by-value relationships, not with by-reference relationships. |
| DICOM PS3.16 2024c - Content Mapping Resource |
|---|