DICOM PS3.16 2024c - Content Mapping Resource

TID 3205 Calibration

The Calibration Template consists of a CONTAINER, with a structure for reporting of the calibration of images used in the analysis.

Table TID 3205. Parameters

Parameter Name

Parameter Usage

$CalibrationPlane

XA Imaging plane


Type:

Extensible

Order:

Significant

Root:

No

Table TID 3205. Calibration

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (122505, DCM, "Calibration")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (111031, DCM, "Image View")

1

U

$CalibrationPlane

3

>

HAS OBS CONTEXT

TEXT

EV (111001, DCM, "Algorithm Name")

1

MC

IF different from Analysis program specified in the invoking Template

4

>

HAS OBS CONTEXT

TEXT

EV (111003, DCM, "Algorithm Version")

1

MC

IF different from Analysis program specified in the invoking Template

5

>

HAS OBS CONTEXT

TEXT

EV (122405, DCM, "Algorithm Manufacturer")

1

MC

IF different from Analysis program specified in the invoking Template

6

>

CONTAINS

CODE

EV (122422, DCM, "Calibration Method")

1

M

DCID 3452 “Calibration Method”

7

>

CONTAINS

CODE

EV (122421, DCM, "Calibration Object")

1

MC

If Row 6 value specifies Calibration Object Used

DCID 3451 “Calibration Object”

8

>

CONTAINS

NUM

EV (122423, DCM, "Calibration Object Size")

1

MC

If Row 6 value specifies Calibration Object Used

DCID 3510 “Catheter Size Unit”

9

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (111026, DCM, "Horizontal Pixel Spacing")

$Units = DT (mm/{pixel}, UCUM, "mm/pixel")

$ImagePurpose = EV (121112, DCM, "Source of Measurement")

10

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (111066, DCM, "Vertical Pixel Spacing")

$Units = DT (mm/{pixel}, UCUM, "mm/pixel")

$ImagePurpose = EV (121112, DCM, "Source of Measurement")

11

>

CONTAINS

IMAGE

1

U


Content Item Descriptions

Row 3

Identifies the Calibration program

Row 4

Identifies the Calibration program version

Row 5

Identifies the Calibration program manufacturer

Row 7

Besides a Sphere and a Catheter, a Distance can be identified as a Calibration Object. In this case a distance measurement of a known dimension of the object is used to calculate the pixel size.

Row 8

The catheter size units is also used to specify the size of other calibration objects (e.g., sphere)

Rows 9, 10

Spacing in the patient body. Point to a single frame containing the image used for calibration if applicable, the actual measurements may be indicated by a SCOORD (see TID 320 “Image or Spatial Coordinates”, Row 3)

Row 11

Secondary Capture image with calibration position. No purpose of reference is specified.

DICOM PS3.16 2024c - Content Mapping Resource