DICOM PS3.16 2024c - Content Mapping Resource

TID 12001 Ultrasound Patient Characteristics

Type:

Extensible

Order:

Significant

Root:

No

Table TID 12001. Ultrasound Patient Characteristics

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (121118, DCM, "Patient Characteristics")

1

M

2

>

CONTAINS

NUM

EV (121033, DCM, "Subject Age")

1

U

UNITS = DCID 7456 “Age Unit”

3

>

CONTAINS

CODE

EV (121032, DCM, "Subject Sex")

1

U

DCID 7455 “Sex”

4

>

CONTAINS

NUM

EV (8302-2, LN, "Patient Height")

1

U

UNITS = EV (cm, UCUM, "cm")

5

>

CONTAINS

NUM

EV (29463-7, LN, "Patient Weight")

1

U

UNITS = EV (kg, UCUM, "kg")

6

>

CONTAINS

NUM

EV (113550, DCM, "Fasting Duration")

1

U

UNITS = EV (h, UCUM, "hours")

7

>

CONTAINS

TEXT

EV (113552, DCM, "Recent Physical Activity")

1

U

8

>

CONTAINS

NUM

EV (8867-4, LN, "Heart Rate")

1

U

UNITS = EV ({H.B.}/min, UCUM, "BPM")

9

>

CONTAINS

NUM

EV (271649006, SCT, "Systolic Blood Pressure")

1

U

UNITS = EV (mm[Hg], UCUM, "mmHg")

10

>

CONTAINS

NUM

EV (271650006, SCT, "Diastolic Blood Pressure")

1

U

UNITS = EV (mm[Hg], UCUM, "mmHg")

11

>

CONTAINS

CODE

EV (260905004, SCT, "Condition")

1-n

U

BCID 12323 “Ultrasound Relevant Patient Condition”

12

>

CONTAINS

INCLUDE

DTID 3923 “BSA-Normalized Ventricular Measurements”

1

U

13

>

CONTAINS

TEXT

EV (121106, DCM, "Comment")

1

U


Content Item Descriptions

Row 6

The number of hours prior to this study that the subject started fasting.

Row 11

Patient conditions present at the time of the procedure that may affect, or are otherwise relevant to, the capture or interpretation of the findings in this report. It is not expected that all known patient conditions will be present here.

Row 13

General comments which may also include relevant patient conditions that would otherwise be recorded in Row 11 if appropriate codes were available.

DICOM PS3.16 2024c - Content Mapping Resource