DICOM PS3.16 2024c - Content Mapping Resource

TID 10054 Procedure Characteristics

This template describes the details of the procedure characteristics (e.g., patient orientation, SID, etc.).

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 10054. Procedure Characteristics

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (130530, DCM, "Procedure Characteristics")

1

M

2

>

CONTAINS

DATETIME

DT (111526, DCM, "DateTime Started")

1

M

3

>

CONTAINS

DATETIME

DT (111527, DCM, "DateTime Ended")

1

M

4

>

CONTAINS

NUM

EV (113832, DCM, "Identification of the X-Ray Source")

1

M

UNITS = EV (1, UCUM, "no units")

5

>

CONTAINS

TEXT

EV (125203, DCM, "Acquisition Protocol")

1

U

6

>

CONTAINS

CODE

EV (113745, DCM, "Patient Table Relationship")

1

U

DCID 21 “Patient Equipment Relationship”

7

>

CONTAINS

CODE

EV (113743, DCM, "Patient Orientation")

1

U

DCID 19 “Patient Orientation”

8

>>

HAS CONCEPT MOD

CODE

EV (113744, DCM, "Patient Orientation Modifier")

1

M

DCID 20 “Patient Orientation Modifier”

9

>

CONTAINS

CODE

EV (123014, DCM, "Target Region")

1

U

DCID 4016 “Anatomic Region for Intra-oral Radiography”

DCID 4026 “Primary Anatomic Structure for Intra-oral and Craniofacial Radiography - Teeth”

DCID 4028 “Craniofacial Anatomic Region”

DCID 4030 “CT, MR and PET Anatomy Imaged”

DCID 4031 “Common Anatomic Region”

10

>>

HAS CONCEPT MOD

CODE

EV (272741003, SCT, "Laterality")

1

UC

IF target region is bilateral

DCID 244 “Laterality”

11

>

CONTAINS

CODE

EV (111635, DCM, "X-Ray Grid")

1-n

U

DCID 10017 “X-Ray Grid”

12

>

CONTAINS

NUM

EV (113750, DCM, "Distance Source to Detector")

1

UC

XOR Row 13

UNITS = EV (mm, UCUM, "mm")

13

>

CONTAINS

TABLE

EV (113750, DCM, "Distance Source to Detector")

1

UC

XOR Row 12

NCOLUMNS = 2

COLUMN 1 = EV (111526, DCM, "DateTime Started")

EV (113750, DCM, "Distance Source to Detector")

COLUMN 2 UNITS = EV (mm, UCUM, "mm")

COLUMN 1 VR = DT

COLUMN 2 VR = FL


Content Item Descriptions

Row 2

The DateTime of the beginning of the time period over which the Content Items are applicable.

Row 3

The DateTime of the end of the time period over which the Content Items are applicable.

Row 4

Identification the X-Ray source. This designation shall not change for a given source throughout the entire RDSR. For systems with multiple X-Ray sources, each source shall be described with a separate instance of this template.

Row 13

In the first table row, the value in COLUMN 1 shall match the value of Row 2 of this template. In the last table row, the value of COLUMN 1 shall not exceed the value of Row 3 of this template.

For all table rows, the value of COLUMN 2 is valid from the time shown in COLUMN 1, until the time shown in COLUMN 1 of the next table row, or in the case of the last table row, until the time shown in Row 3 of this template.

DICOM PS3.16 2024c - Content Mapping Resource