| DICOM PS3.16 2024c - Content Mapping Resource |
|---|
This template describes the details of the attenuator positions.
Table TID 10052. Attenuator Position
|
IF an X-Ray Attenuator is described in TID 10047 AND the position is known |
||||||||
|
The DateTime of the beginning of the time period over which the Content Items are applicable. |
|
|
The DateTime of the end of the time period over which the Content Items are applicable. |
|
|
Provide the model and position of each X-Ray attenuator described in TID 10047, if known. |
|
|
Each X-Ray attenuator model and position description includes the identification of the X-Ray attenuator, which shall match the identification used in TID 10047. |
|
|
Reference to a three-dimensional model of the X-Ray attenuator |
|
|
A 4-by-4 matrix of dimensionless numbers of the form defined in Section C.20.2.1.1 “Frame of Reference Transformation Matrix” in PS3.3 . The matrix describes the rigid transformation matrix (including translation and rotation) that transforms the reference coordinate system for the X-Ray attenuator to the RDSR reference coordinate system. The translation described by this matrix indicates the position and orientation of the X-Ray attenuator within the RDSR reference coordinate system. The RCS for the X-Ray attenuator shall be a right-handed Cartesian coordinate system. If the model data in Rows 6, 7, 8 was not originally in a right-handed Cartesian coordinate system, it must transformed prior to inclusion in this template. |
| DICOM PS3.16 2024c - Content Mapping Resource |
|---|