1
|
|
|
CONTAINER
|
EV (130511, DCM, "Radiation Technique")
|
1
|
M
|
|
|
|
2
|
>
|
CONTAINS
|
DATETIME
|
DT (111526, DCM, "DateTime Started")
|
1
|
M
|
|
|
|
3
|
>
|
CONTAINS
|
DATETIME
|
DT (111527, DCM, "DateTime Ended")
|
1
|
M
|
|
|
|
4
|
>
|
CONTAINS
|
TEXT
|
EV (113832, DCM, "Identification of the X-Ray Source")
|
1
|
M
|
|
|
|
5
|
>
|
CONTAINS
|
NUM
|
EV (111634, DCM, "Half Value Layer")
|
1
|
UC
|
XOR Row 7
|
UNITS = EV (mm, UCUM, "mm")
|
|
6
|
>>
|
HAS PROPERTIES
|
CODE
|
EV (130510, DCM, "Reported Value Type")
|
1
|
U
|
|
DCID 10072 “Reported Value Type”
|
|
7
|
>
|
CONTAINS
|
TABLE
|
EV (111634, DCM, "Half Value Layer")
|
1
|
UC
|
XOR Row 5
|
NCOLUMNS = 2
COLUMN 1 = EV (111526, DCM, "DateTime Started")
COLUMN 2 = EV (111634, DCM, "Half Value Layer")
COLUMN 2 UNITS = EV (mm, UCUM, "mm")
COLUMN 1 VR = DT
COLUMN 2 VR = FL
|
|
8
|
>>
|
HAS PROPERTIES
|
CODE
|
EV (130510, DCM, "Reported Value Type")
|
1
|
U
|
|
DCID 10072 “Reported Value Type”
|
|
9
|
>
|
CONTAINS
|
NUM
|
EV (113733, DCM, "KVP")
|
1
|
MC
|
XOR Row 11
|
UNITS = EV (kV, UCUM, "kV")
|
|
10
|
>>
|
HAS PROPERTIES
|
CODE
|
EV (130510, DCM, "Reported Value Type")
|
1
|
U
|
|
DCID 10072 “Reported Value Type”
|
|
11
|
>
|
CONTAINS
|
TABLE
|
EV (113733, DCM, "KVP")
|
1
|
MC
|
XOR Row 9
|
NCOLUMNS = 2
COLUMN 1 = EV (111526, DCM, "DateTime Started")
COLUMN 2 = EV (113733, DCM, "KVP")
COLUMN 2 UNITS = EV (kV, UCUM, "kV")
COLUMN 1 VR = DT
COLUMN 2 VR = FL
|
|
12
|
>>
|
HAS PROPERTIES
|
CODE
|
EV (130510, DCM, "Reported Value Type")
|
1
|
U
|
|
DCID 10072 “Reported Value Type”
|
|
13
|
>
|
CONTAINS
|
NUM
|
EV (113734, DCM, "X-Ray Tube Current")
|
1
|
MC
|
XOR Row 15
|
UNITS = EV (mA, UCUM, "mA")
|
|
14
|
>>
|
HAS PROPERTIES
|
CODE
|
EV (130510, DCM, "Reported Value Type")
|
1
|
U
|
|
DCID 10072 “Reported Value Type”
|
|
15
|
>
|
CONTAINS
|
TABLE
|
EV (113734, DCM, "X-Ray Tube Current")
|
1
|
MC
|
XOR Row 13
|
NCOLUMNS = 2
COLUMN 1 = EV (111526, DCM, "DateTime Started")
COLUMN 2 = EV (113734, DCM, "X-Ray Tube Current")
COLUMN 2 UNITS = EV (mA, UCUM, "mA")
COLUMN 1 VR = DT
COLUMN 2 VR = FL
|
|
16
|
>>
|
HAS PROPERTIES
|
CODE
|
EV (130510, DCM, "Reported Value Type")
|
1
|
U
|
|
DCID 10072 “Reported Value Type”
|
|
17
|
>
|
CONTAINS
|
NUM
|
EV (113791, DCM, "Pulse Rate")
|
1
|
UC
|
XOR Row 19
|
UNITS = EV ({pulse}/s, UCUM, "pulse/s")
|
|
18
|
>>
|
HAS PROPERTIES
|
CODE
|
EV (130510, DCM, "Reported Value Type")
|
1
|
U
|
|
DCID 10072 “Reported Value Type”
|
|
19
|
>
|
CONTAINS
|
TABLE
|
EV (113791, DCM, "Pulse Rate")
|
1
|
UC
|
XOR Row 17
|
NCOLUMNS = 2
COLUMN 1 = EV (111526, DCM, "DateTime Started")
COLUMN 2 = EV (113791, DCM, "Pulse Rate")
COLUMN 2 UNITS = EV ({pulse}/s, UCUM, "pulse/s")
COLUMN 1 VR = DT
COLUMN 2 VR = FL
|
|
20
|
>>
|
HAS PROPERTIES
|
CODE
|
EV (130510, DCM, "Reported Value Type")
|
1
|
U
|
|
DCID 10072 “Reported Value Type”
|
|
21
|
>
|
CONTAINS
|
NUM
|
EV (113793, DCM, "Pulse Width")
|
1
|
UC
|
XOR Row 23
|
UNITS = EV (ms, UCUM, "ms")
|
|
22
|
>>
|
HAS PROPERTIES
|
CODE
|
EV (130510, DCM, "Reported Value Type")
|
1
|
U
|
|
DCID 10072 “Reported Value Type”
|
|
23
|
>
|
CONTAINS
|
TABLE
|
EV (113793, DCM, "Pulse Width")
|
1
|
UC
|
XOR Row 21
|
NCOLUMNS = 2
COLUMN 1 = EV (111526, DCM, "DateTime Started")
COLUMN 2 = EV (113793, DCM, "Pulse Width")
COLUMN 2 UNITS = EV (ms, UCUM, "ms")
COLUMN 1 VR = DT
COLUMN 2 VR = FL
|
|
24
|
>>
|
HAS PROPERTIES
|
CODE
|
EV (130510, DCM, "Reported Value Type")
|
1
|
U
|
|
DCID 10072 “Reported Value Type”
|