DICOM PS3.16 2024c - Content Mapping Resource

TID 10003A Irradiation Event X-Ray Detector Data

This Template contains data that is expected to be available to the X-Ray detector or plate reader component of the equipment.

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 10003A. Irradiation Event X-Ray Detector Data

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

NUM

EV (113845, DCM, "Exposure Index")

1

MC

IF the value is displayable to the X-Ray system operator.

UNITS = EV (1, UCUM, "no units")

2

NUM

EV (113846, DCM, "Target Exposure Index")

1

MC

IF the value is displayable to the X-Ray system operator.

UNITS = EV (1, UCUM, "no units")

3

NUM

EV (113847, DCM, "Deviation Index")

1

MC

IF the value is displayable to the X-Ray system operator.

UNITS = EV (1, UCUM, "no units")

4

INCLUDE

DTID 1021 “Device Participant”

1

U

$DeviceProcedureRole = EV (113942, DCM, "X-Ray Reading Device")

5

IMAGE

EV (113795, DCM, "Acquired Image")

1-n

MC

IFF Image Object is created for this irradiation event


Content Item Descriptions

Row 4

The device that read the detector of this Irradiation Event. E.g., the CR Plate Reader.

Row 5

Reference to Image instances created during this event, if any. The UID reference(s) provided here shall be the values at the time the images were initially created. (Note that image UIDs may be changed as the images are managed over a long term.)

DICOM PS3.16 2024c - Content Mapping Resource