Event
|
EventID
|
M
|
EV (110102, DCM, "Begin Transferring DICOM Instances")
|
|
EventActionCode
|
M
|
Shall be: E = Execute
|
|
EventDateTime
|
M
|
not specialized
|
|
EventOutcomeIndicator
|
M
|
not specialized
|
|
EventTypeCode
|
U
|
not specialized
|
|
Active Participant:
Process Sending the Data (1)
|
UserID
|
M
|
The identity of the process sending the data.
|
|
AlternativeUserID
|
U
|
not specialized
|
|
UserName
|
U
|
not specialized
|
|
UserIsRequestor
|
M
|
not specialized
|
|
RoleIDCode
|
M
|
EV (110153, DCM, "Source Role ID")
|
|
NetworkAccessPointTypeCode
|
U
|
not specialized
|
|
NetworkAccessPointID
|
U
|
not specialized
|
|
Active Participant:
Process receiving the data (1)
|
UserID
|
M
|
The identity of the process receiving the data.
|
|
AlternativeUserID
|
U
|
not specialized
|
|
UserName
|
U
|
not specialized
|
|
UserIsRequestor
|
M
|
not specialized
|
|
RoleIDCode
|
M
|
EV (110152, DCM, "Destination Role ID")
|
|
NetworkAccessPointTypeCode
|
U
|
not specialized
|
|
NetworkAccessPointID
|
U
|
not specialized
|
|
Active Participant:
Other Participants (0..N)
|
UserID
|
M
|
The identity of any other participants that might be involved and known, especially third parties that are the requestor
|
|
AlternativeUserID
|
U
|
not specialized
|
|
UserName
|
U
|
not specialized
|
|
UserIsRequestor
|
M
|
not specialized
|
|
RoleIDCode
|
U
|
not specialized
|
|
NetworkAccessPointTypeCode
|
U
|
not specialized
|
|
NetworkAccessPointID
|
U
|
not specialized
|
|
Participating Object:
Studies being transferred (1..N)
|
ParticipantObjectTypeCode
|
M
|
Shall be: 2 = System Object
|
|
ParticipantObjectTypeCodeRole
|
M
|
Shall be: 3 = Report
|
|
ParticipantObjectDataLifeCycle
|
U
|
not specialized
|
|
ParticipantObjectIDTypeCode
|
M
|
EV (110180, DCM, "Study Instance UID")
|
|
ParticipantObjectSensitivity
|
U
|
not specialized
|
|
ParticipantObjectID
|
M
|
The Study Instance UID
|
|
ParticipantObjectName
|
U
|
not specialized
|
|
ParticipantObjectQuery
|
U
|
not specialized
|
|
ParticipantObjectDetail
|
U
|
Element "ContainsSOPClass" with one or more SOP Class UID values
|
|
ParticipantObjectDescription
|
U
|
not specialized
|
|
SOPClass
|
MC
|
not specialized
|
|
Accession
|
U
|
not specialized
|
|
NumberOfInstances
|
U
|
not specialized
|
|
Instances
|
U
|
not specialized
|
|
Encrypted
|
U
|
not specialized
|
|
Anonymized
|
U
|
not specialized
|
|
Participating Object:
Patient (1)
|
ParticipantObjectTypeCode
|
M
|
Shall be: 1 = Person
|
|
ParticipantObjectTypeCodeRole
|
M
|
Shall be: 1 = Patient
|
|
ParticipantObjectDataLifeCycle
|
U
|
not specialized
|
|
ParticipantObjectIDTypeCode
|
M
|
Shall be: 2 = Patient Number
|
|
ParticipantObjectSensitivity
|
U
|
not specialized
|
|
ParticipantObjectID
|
M
|
The patient ID
|
|
ParticipantObjectName
|
U
|
The patient name
|
|
ParticipantObjectQuery
|
U
|
not specialized
|
|
ParticipantObjectDetail
|
U
|
not specialized
|
|
ParticipantObjectDescription
|
U
|
not specialized
|