| DICOM PS3.11 2024c - Media Storage Application Profiles |
|---|
Application Profiles in this class, STD-US, shall support the appropriate Information Object Definitions (IOD) and Transfer Syntaxes for the Media Storage SOP Class in the following table. In the role of FS-Updater or FS-Creator the application can choose one of the three possible transfer Syntaxes to create an IOD. In the role of FS-Reader an application shall support all transfer Syntaxes defined for the STD-US application profile.
Table C.3-1. Ultrasound SOP Classes and Transfer Syntaxes
|
1.2.840.10008.1.2.1 (see Section 8.6 in PS3.10 ) |
|||
|
JPEG Lossy, Baseline Sequential with Huffman Coding (Process 1) |
|||
|
JPEG Lossy, Baseline Sequential with Huffman Coding (Process 1) |
The STD-US application profile requires that all ultrasound image objects only be stored using the values described in PS3.3 US Image Module and the specializations used for the Ultrasound Single and Multi-Frame IODs.
In the role of FS-Updater or FS-Creator the application can choose any of the supported Photometric Interpretations described in PS3.3 US Image Module to create an IOD. In the role of FS-Reader, an application shall support all Photometric Interpretations described in PS3.3 US Image Module.
Table C.3-2 describes restrictions on the use of various Transfer Syntaxes with the supported Photometric Interpretations for both single and multi-frame images.
| DICOM PS3.11 2024c - Media Storage Application Profiles |
|---|