| DICOM PS3.11 2024c - Media Storage Application Profiles |
|---|
This section defines other parameters common to all specific Application Profiles in the STD-XABC class that need to be specified in order to ensure interoperable media interchange.
The attributes listed in Table A.3-3 used within the X-Ray Angiographic Image files shall take the values specified.
When creating or updating a File-set, Rows or Columns shall not exceed a value of 512. When reading a File-set, an FSR or FSU shall accept a value of at least 512 for Rows or Columns.
Overlay data, if present, shall be encoded in Overlay Data (60XX,3000).
| DICOM PS3.11 2024c - Media Storage Application Profiles |
|---|