DICOM PS3.7 2024c - Message Exchange

C.4 Warning Status Class

Statuses in this Status Class convey that the operation/notification has completed but an error was detected. The semantics and behavior of these Statuses are defined in PS3.4.

C.4.1 Warning

Status Field

Tag

VR

VM

Description of Field

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field is Service Class specific and defined in PS3.4.

Offending Element

(0000,0901)

AT

1-n

This optional field contains a list of the elements in which the error was detected.

Error Comment

(0000,0902)

LO

1

This optional field contains an application-specific text description of the error detected.

Note

Only characters from the Default Character Repertoire may be used.

DICOM PS3.7 2024c - Message Exchange