| DICOM PS3.7 2024c - Message Exchange |
|---|
The information necessary for the C-GET request and indication DIMSE-C primitives are conveyed in the C-GET-RQ Message. The information necessary for the C-GET response and confirmation DIMSE-C primitives are conveyed in the C-GET-RSP Message. The information necessary for the C-GET Cancel Request and Cancel Indication primitives are conveyed in the C-CANCEL-GET-RQ Message.
The C-GET-RQ Message contains fields as defined in Table 9.3-6. Each field shall conform to DICOM encoding and Value Representation as defined in PS3.5. Fields are required as specified in the C-GET service definition unless otherwise noted in Table 9.3-6. Fields not specified in the C-GET service definition but present in Table 9.3-6 are required by the DIMSE-C protocol.
Table 9.3-6. C-GET-RQ Message Fields
|
The even number of bytes from the end of the value field to the beginning of the next group. |
||||
|
This field distinguishes the DIMSE-C operation conveyed by this Message. The value of this field shall be set to 0010H for the C-GET-RQ Message. |
||||
|
Implementation-specific value that distinguishes this Message from other Messages. |
||||
|
This field indicates that a Data Set is present in the Message. It shall be set to any value other than 0101H (Null). |
||||
|
A Data Set that encodes attributes providing status information about the C-GET operation. See Section 9.1.3.1.5. |
| DICOM PS3.7 2024c - Message Exchange |
|---|