| DICOM PS3.4 2024c - Service Class Specifications |
|---|
The DICOM AE's Conformance Statement shall be formatted as defined in PS3.2.
The SCU shall document in its Conformance Statement the behavior and actions that cause the SCU to generate an N-ACTION-RQ primitive.
The SCU shall document how it obtains the Patient ID (0010,0020) or Admission ID (0038,0010) Attribute (e.g., through a Modality Worklist query, bar-code scan, manual entry, etc.).
The SCU shall document the behavior and actions performed when a success or failure status is received.
| DICOM PS3.4 2024c - Service Class Specifications |
|---|