| DICOM PS3.4 2024c - Service Class Specifications |
|---|
The DIMSE Services that are applicable to the IOD are shown in Table H.4.11.2-1.
The meaning of the Usage SCU/SCP is described in Section H.2.4.
This Section describes the behavior of the DIMSE Service that are specific for this IOD. The general behavior of the DIMSE Services is specified in PS3.7.
The N-GET is used to retrieve an instance of the Printer Configuration Retrieval SOP Class.
The Attributes that are retrieved are shown in Table H.4-26.
Table H.4-26. N-GET Attributes
The meaning of the Usage SCU/SCP is described in Section H.2.4.
The SCU uses the N-GET to request the SCP to get a Printer Configuration Retrieval SOP Instance. The SCU shall specify in the N-GET request primitive the UID of the SOP Instance to be retrieved.
The SCP shall return the values for the specified Attributes of the specified SOP Instance.
The SCP shall return the Status Code of the requested SOP Instance retrieval.
A Failure Status Code shall indicate that the SCP has not retrieved the SOP Instance.
There are no specific Status Codes. See PS3.7 for response Status Codes.
| DICOM PS3.4 2024c - Service Class Specifications |
|---|