| DICOM PS3.4 2024c - Service Class Specifications |
|---|
The implementation Conformance Statement of these SOP Classes shall follow PS3.2.
The SCU Conformance Statement shall specify the following items:
The SCP Conformance Statement shall specify the following items:
minimum and maximum number of printable pixel matrix per supported film size
list of supported optional SOP Class Attributes and DICOM Message Service Elements
for each supported Attribute (mandatory and optional Attribute):
Status Code (Failure or Warning) if SCU supplies a value that is out of range
for each supported DIMSE Service, the SCP behavior for all specific Status Codes
description of each supported custom Image Display Format (2010,0010) e.g., position and dimensions of each composing image box, numbering scheme of the image positions
description of each supported Annotation Display Format ID (2010,0030) e.g., position and dimensions of annotation box, font, number of characters
description of each supported configuration table (e.g., identification, content)
if the SCP supports N-ACTION for the Film Session SOP Class then the SCP shall specify the maximum number of collated films
in the case of grayscale printers that print color images, the behavior of printing color images
if cropping of images is supported, the algorithm for removing rows and columns from the image
| DICOM PS3.4 2024c - Service Class Specifications |
|---|