| DICOM PS3.4 2024c - Service Class Specifications |
|---|
The following issues shall be documented in the Conformance Statement of any implementation claiming conformance to the Storage Service Class as an SCP:
The Level of Storage Support, as defined by Section B.4.1, shall be stated.
The Level of Digital Signature Support, as defined by Section B.4.1, shall be stated.
The optional Attributes that will be discarded (if any) shall be listed for each IOD supported.
The mechanisms by which additional SOP Classes are dynamically supported, as defined by Section B.4.1.2, shall be stated.
The Conformance Statement shall document the policies concerning the Attribute Lossy Image Compression (0028,2110).
The behavior of the SCP in the case of a successful C-STORE operation shall be described. This includes the following:
The meaning of each case of an unsuccessful C-STORE response status shall be described, as well as appropriate recovery action.
The meaning of each case of a warning C-STORE response status shall be described, as well as appropriate action.
If the SCP performs coercion on any Attributes, this shall be stated, and the conditions under which it may occur shall be described.
| DICOM PS3.4 2024c - Service Class Specifications |
|---|