| DICOM PS3.3 2024c - Information Object Definitions |
|---|
The RT ROI Observations Module specifies the identification and interpretation of an ROI specified in the Structure Set Module and ROI Contour Module.
Table C.8-44. RT ROI Observations Module Attributes
|
Sequence of observations related to ROIs defined in the Structure Set ROI Sequence (3006,0020) of the Structure Set Module. |
|||
|
Identification number of the Observation. The value of Observation Number (3006,0082) shall be unique within the RT ROI Observations Sequence (3006,0080). |
|||
|
Uniquely identifies the referenced ROI described in the Structure Set ROI Sequence (3006,0020). |
|||
|
Sequence of significantly related ROIs, e.g., CTVs contained within a PTV. |
|||
|
Uniquely identifies the related ROI described in the Structure Set ROI Sequence (3006,0020). |
|||
|
Relationship of referenced ROI with respect to referencing ROI. |
|||
|
>Include Table 10-7b “Multiple Site General Anatomy Optional Macro Attributes” |
May not be necessary if the anatomy is implicit in the RT ROI Identification Code Sequence (3006,0086). More than one item in Anatomic Region Sequence (0008,2218) may be used when a region of interest spans multiple anatomical locations and there is not a single pre-coordinated code describing the combination of locations. There is no requirement that the multiple locations be contiguous. |
||
|
Sequence defining the general category of the property this ROI represents. Only a single Item is permitted in this Sequence. See Note 1. |
|||
|
Sequence containing Code used to identify ROI. Only a single Item is permitted in this Sequence. See Note 2. |
|||
|
See Section C.8.8.8.3 RT ROI Identification Code Sequence and Therapeutic Role Type Code Sequence. |
|||
|
Sequence defining the modifier of the property type of this segment. |
|||
|
>>>Include Table 8.8-1 “Code Sequence Macro Attributes” |
|||
|
The specific property type of the therapeutic role of this ROI. |
|||
|
See Section C.8.8.8.3 RT ROI Identification Code Sequence and Therapeutic Role Type Code Sequence. |
|||
|
Type of ROI. See Section C.8.8.8.1. Defined Terms:
|
|||
|
Person or device performing the interpretation. Required if ROI Creator Sequence (3006,004D) is present in the Structure Set Module, and the person or device performing the interpretation differs from ROI Creator Sequence (3006,004D). May be present otherwise. |
|||
|
>>Include Table C.17-3b “Identified Person or Device Macro Attributes” |
Organizational Role Code Sequence (0044,010A) DCID 9555 “Radiotherapy Treatment Planning Person Role”. |
||
|
Sequence describing physical properties associated with current ROI interpretation. |
|||
|
Physical property specified by ROI Physical Property Value (3006,00B4). |
|||
|
The elemental composition of the ROI and the atomic mass fraction of the elements in the ROI. Required if ROI Physical Property (3006,00B2) equals ELEM_FRACTION. |
|||
|
The atomic number of the element for which the ROI Elemental Composition Sequence (3006,00B6) is present. |
|||
|
The fractional weight of the element for which the ROI Elemental Composition Sequence (3006,00B6) is present. The sum of all ROI Elemental Composition Atomic Mass Fractions (3006,00B8) within the ROI Elemental Composition Sequence (3006,00B6) shall equal 1.0 within acceptable limits of floating point precision. |
|||
This Attribute allows preserving information by copying the content of Segmented Property Category Code Sequence (0062,0003) in case a Segmentation object is re-encoded as an RT Structure Set or vice-versa.
In case of re-encoding a Segmentation object as an RT Structure Set or vice-versa it is suggested, that the Segmented Property Type Code Sequence (0062,000F) is mapped to RT ROI Identification Code Sequence (3006,0086).
RT ROI Interpreted Type (3006,00A4) shall describe the class of ROI (e.g., CTV, PTV).
In addition to providing a Defined Term in RT ROI Interpreted Type (3006,00A4), it is recommended that coded information be included; see Section C.8.8.8.3 RT ROI Identification Code Sequence and Therapeutic Role Type Code Sequence.
| DICOM PS3.3 2024c - Information Object Definitions |
|---|