| DICOM PS3.3 2024c - Information Object Definitions |
|---|
A structure set defines a set of areas of significance. Each area can be associated with a Frame of Reference and zero or more images. Information that can be transferred with each region of interest (ROI) includes geometrical and display parameters, and generation technique.
Table C.8-41. Structure Set Module Attributes
|
Date at which the content of the Structure Set was last modified. |
|||
|
Time at which the content of the Structure Set was last modified. |
|||
|
Sequence describing Frames of Reference in which the ROIs are defined. One or more Items are permitted in this Sequence. See Section C.8.8.5.1. |
|||
|
Uniquely identifies Frame of Reference within Structure Set. |
|||
|
Sequence of Studies containing Series to be referenced. One or more Items are permitted in this Sequence. See Section C.8.8.5.4. |
|||
|
>>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
|
Sequence describing Series of images within the referenced Study that are used in defining the Structure Set. |
|||
|
Sequence of Items describing images in a given Series used in defining the Structure Set (typically CT or MR images). |
|||
|
>>>>Include Table 10-3 “Image SOP Instance Reference Macro Attributes” |
|||
|
Information about Image Series that are sources of ROIs in this Instance. This also includes Series present in the RT Referenced Series Sequence (3006,0014). The Image Series may have been the source of the definition, but the ROI has been resampled to the Image Series referenced in the Referenced Frame of Reference Sequence (3006,0010). |
|||
|
Type of device, process or method that originally acquired or produced the data used to create the Instances in this Series. |
|||
|
Identification number of the ROI. The value of ROI Number (3006,0022) shall be unique within the Structure Set in which it is created. |
|||
|
Identifies the Conceptual Volume which is represented by this ROI. |
|||
|
>>Include Table 10.33-1 “Conceptual Volume Macro Attributes”. |
|||
|
Uniquely identifies Frame of Reference in which ROI is defined, specified by Frame of Reference UID (0020,0052) in Referenced Frame of Reference Sequence (3006,0010). |
|||
|
The protocol(s) selected by the RT Physician. One or more Items are permitted in this Sequence. See Section C.36.5.1.2. |
|||
|
>>Include Table C.17-3b “Identified Person or Device Macro Attributes” |
Organizational Role Code Sequence (0044,010A) DCID 9555 “Radiotherapy Treatment Planning Person Role”. |
||
|
>>Include Table 10-19 “Algorithm Identification Macro Attributes”. |
|||
|
A coded description of how this ROI was derived. One or more Items are permitted in this Sequence. See Section C.8.8.5.3 for further explanation. |
|||
|
>>Include Table 8.8-1 “Code Sequence Macro Attributes”. |
Enumerated Value (113085, DCM, "Spatial resampling"). |
||
|
Instances containing the source of the ROI Contour information. |
|||
|
>>Include Table 10-11 “SOP Instance Reference Macro Attributes”. |
|||
|
The value of Segment Number (0062,0004) in the referenced SOP Instance that identifies the segment that is the source of the ROI Contour information. Required if Referenced SOP Class UID (0008,1150) is Segmentation Storage ("1.2.840.10008.5.1.4.1.1.66.4"). |
|||
|
The value of Fiducials UID (0070,031A) in the referenced SOP Instance that identifies the fiducial that is the source of the ROI Contour information. Required if Referenced SOP Class UID (0008,1150) is Spatial Fiducials Storage ("1.2.840.10008.5.1.4.1.1.66.2"). |
|||
|
The Structure Set that has been used to derive the current Structure Set. |
|||
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
The Referenced Frame of Reference Sequence (3006,0010) describes a set of frames of reference in which some or all of the ROIs are expressed. Since the Referenced Frame of Reference UID (3006,0024) is required for each ROI, each Frame of Reference used to express the coordinates of an ROI shall be listed in the Referenced Frame of Reference Sequence (3006,0010) once and only once.
As an example, a set of ROIs defined using a single image Series would list the image Series in a single Referenced Frame of Reference Sequence (3006,0010) Item, providing the UID for this referenced Frame of Reference (obtained from the source images), and listing all pertinent images in the Contour Image Sequence (3006,0016).
| DICOM PS3.3 2024c - Information Object Definitions |
|---|