| DICOM PS3.3 2024c - Information Object Definitions |
|---|
The RT Dose Module is used to convey 2D or 3D radiation dose data generated from treatment planning systems or similar devices. The Attributes defined within the Module support dose for a single radiation beam (potentially comprised of multiple segments, as delivered in a dynamic treatment) or a group of beams comprising either a fraction group (see Section C.8.8.13) or a complete treatment plan (potentially the sum of multiple fraction groups).
The RT Dose Module provides the mechanism to transmit a 3D array of dose data as a set of 2D dose planes that may or may not be related to CT or MR image planes. This mechanism works via the DICOM Multi-frame Module that is required if multi-frame pixel data are present.
Table C.8-39. RT Dose Module Attributes
|
Number of samples (planes) in this image. See Section C.8.8.3.4.1 for specialization. Required if Pixel Data (7FE0,0010) is present. |
|||
|
Specifies the intended interpretation of the pixel data. See Section C.8.8.3.4.2 for specialization. Required if Pixel Data (7FE0,0010) is present. |
|||
|
Number of bits allocated for each pixel sample. Each sample shall have the same number of bits allocated. See Section C.8.8.3.4.3 for specialization. Required if Pixel Data (7FE0,0010) is present. |
|||
|
Number of bits stored for each pixel sample. Each sample shall have the same number of bits stored. See Section C.8.8.3.4.4 for specialization. Required if Pixel Data (7FE0,0010) is present. |
|||
|
Most significant bit for each pixel sample. Each sample shall have the same high bit. See Section C.8.8.3.4.5 for specialization. Required if Pixel Data (7FE0,0010) is present. |
|||
|
Data representation of the pixel samples. Each sample shall have the same pixel representation. See Section C.8.8.3.4.6 for specialization. Required if Pixel Data (7FE0,0010) is present. |
|||
|
Defined Terms: See Section C.8.8.3.6. |
|||
|
The use of transformation in the calculation of the combined dose. Defined Terms:
|
|||
|
A reference to a Spatial Registration SOP Instance or a Deformable Spatial Registration SOP Instance, which defines the transformation used to transform the dose. Required, if Spatial Transform of Dose (3004,0005) is provided and has a value of RIGID or NON_RIGID. |
|||
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
|
User-defined label for dose data. See Section 10.32.2.1. |
|||
|
Coordinates (x, y, z) of normalization point in the patient-Based Coordinate System described in Section C.7.6.2.1.1 (mm). See Section C.8.8.3.1. |
|||
|
Defined Terms:
|
|||
|
Sequence describing RT Plan associated with dose. Required if Dose Summation Type (3004,000A) is PLAN, MULTI_PLAN, FRACTION, BEAM, BRACHY, FRACTION_SESSION, BEAM_SESSION, BRACHY_SESSION or CONTROL_POINT. May be present if Dose Summation Type (3004,000A) is PLAN_OVERVIEW. Only a single Item shall be included in this Sequence, unless Dose Summation Type (3004,000A) is MULTI_PLAN, in which case two or more Items shall be included in this Sequence. See Note 1. |
|||
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
|
The value of Plan Overview Index (300C,0117) from the Plan Overview Sequence (300C,0116) to which this RT Plan corresponds. Shall be unique, i.e., not be duplicated within another Item of this Referenced RT Plan Sequence (300C,0002). |
|||
|
Sequence of one Fraction Group containing beams or brachy application setups contributing to dose. Required if Dose Summation Type (3004,000A) is FRACTION, BEAM, BRACHY, FRACTION_SESSION, BEAM_SESSION, BRACHY_SESSION or CONTROL_POINT. Only a single Item shall be included in this Sequence. See Note 1. |
|||
|
Uniquely identifies Fraction Group specified by Fraction Group Number (300A,0071) in Fraction Group Sequence of RT Fraction Scheme Module within RT Plan referenced in Referenced RT Plan Sequence (300C,0002). |
|||
|
Sequence of Beams in current Fraction Group contributing to dose. Required if Dose Summation Type (3004,000A) is BEAM, BEAM_SESSION or CONTROL_POINT. |
|||
|
Uniquely identifies Beam specified by Beam Number (300A,00C0) in Beam Sequence (300A,00B0) of RT Beams Module within RT Plan referenced in Referenced RT Plan Sequence (300C,0002) or in Ion Beam Sequence (300A,03A2) of RT Ion Beams Module within RT Ion Plan referenced in Referenced RT Plan Sequence (300C,0002). |
|||
|
Sequence defining the Control Points in current Beam contributing to dose. Required if Dose Summation Type (3004,000A) is CONTROL_POINT. |
|||
|
Identifies Control Point specified by Control Point Index (300A,0112) within Beam referenced by Referenced Beam Number (300C,0006). This is the first of the two Control Points from which the Dose contribution to the Control Point can be calculated. |
|||
|
Identifies Control Point specified by Control Point Index (300A,0112) within Beam referenced by Referenced Beam Number (300C,0006). This is the second of the two Control Points from which the Dose contribution to the Control Point can be calculated. The Control Point Index (300A,0112) referenced by Referenced Stop Control Point Index (300C,00F6) shall be the Control Point Index (300A,0112) immediately following the Control Point Index (300A,0112) referenced by Referenced Start Control Point Index (300C,00F4) within the Referenced Beam Number (300C,0006). |
|||
|
Sequence of Brachy Application Setups in current Fraction Group contributing to dose. Required if Dose Summation Type (3004,000A) is BRACHY or BRACHY_SESSION. One or more Items shall be included in this Sequence. |
|||
|
Uniquely identifies Brachy Application Setup specified by Brachy Application Setup Number (300A,0234) in Brachy Application Setup Sequence (300A,0230) of RT Brachy Application Setups Module within RT Plan referenced in Referenced RT Plan Sequence (300C,0002). |
|||
|
Sequence describing RT Beams Treatment Record associated with dose. |
|||
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
|
Sequence of Beams contributing to dose. One or more Items shall be included in this Sequence. Required, if the dose does not apply to the complete RT Beams Treatment Record referenced in the Referenced Treatment Record Sequence (3008,0030). |
|||
|
Uniquely identifies Beam specified by Referenced Beam Number (300C,0006) in Treatment Session Beam Sequence (3008,0020) of RT Beams Session Record Module within RT Beams Treatment Record referenced in the Referenced Treatment Record Sequence (3008,0030) or in Treatment Session Ion Beam Sequence (3008,0021) of RT Ion Beams Session Record Module within RT Ion Beams Treatment Record referenced in the Referenced Treatment Record Sequence (3008,0030). |
|||
|
An array that contains the dose image plane offsets (in mm) of the dose image frames in a multi-frame dose. Required if multi-frame pixel data are present and Frame Increment Pointer (0028,0009) points to Grid Frame Offset Vector (3004,000C). See Section C.8.8.3.2. |
|||
|
Scaling factor that when multiplied by the dose grid data found in Pixel Data (7FE0,0010) Attribute of the Image Pixel Module, yields grid doses in the dose units as specified by Dose Units (3004,0002). |
|||
|
Specifies a list of patient heterogeneity characteristics used for calculating dose. This Attribute shall be multi-valued if beams used to compute the dose have differing correction techniques. |
|||
|
A coded description of how this dose was derived from other RT Dose and/or RT Plan objects. One or more Items are permitted in this Sequence. More than one Item indicates that successive derivation steps have been applied. |
|||
|
The set of SOP Instances used to derive this RT Dose SOP Instance from other RT Dose and/or RT Plan objects. |
|||
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
|
Code describing the purpose of the reference to the Instance(s). |
|||
|
Parameters providing an overview of the plan used to create this RT Dose Instance. Required if Dose Summation Type (3004,000A) is PLAN_OVERVIEW. May be present if Dose Summation Type (3004,000A) is PLAN, MULTI_PLAN or RECORD. One or more Items shall be included in this Sequence if Dose Summation Type (3004,000A) is PLAN_OVERVIEW. Only one Item shall be included in this Sequence if Dose Summation Type (3004,000A) is PLAN or RECORD. Two or more Items shall be included in this Sequence if Dose Summation Type (3004,000A) is MULTI_PLAN. |
|||
|
Number of fractions of the plan included in this RT Dose Instance. Required if Dose Summation Type (3004,000A) is PLAN_OVERVIEW, PLAN or MULTI_PLAN. If Dose Summation Type is PLAN or MULTI_PLAN the Number of Fractions Included (300C,0119) shall equal the Number of Fractions Planned (300A,0078) of the referenced RT Plan Instance. |
|||
|
Fraction Number of the fraction included in this RT Dose Instance. |
|||
|
>>>Include Table 8.8-1 “Code Sequence Macro Attributes” |
|||
|
Label identifying the Prescription Overview. For example, this may be the name of the related target ROI or the description of a Dose Reference. Required if Prescription Overview Sequence (300C,0114) has more than one Item. May be present otherwise. |
|||
|
Prescribed total dose in Gy for all fractions for the dose type defined in Dose Type (3004,0004). |
|||
|
ROI for which the prescription parameters in this Sequence apply, specified by ROI Number (3006,0022) in the Instance referenced by Referenced Structure Set Sequence (300C,0060). |
|||
|
Structure Set containing structures that were used to calculate the RT Dose. Required if Referenced Image Sequence (0008,1140) is not present. |
|||
|
>>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
|
Images used to calculate the RT Dose. Required if Referenced Structure Set Sequence (300C,0060) is not present. |
|||
|
>>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
In order to prevent misrepresentation of the dose summation:
If the Dose Summation Type (3004,000A) is PLAN, then only a single Instance of RT Plan is referenced and the dose will be for the entire plan (i.e., it is not viable to combine only certain fraction groups of different plans).
If the Dose Summation Type (3004,000A) is MULTI_PLAN, then 2 or more Instances of RT Plan may be referenced. As above, each reference will be for the entire plan.
If the Dose Summation Type (3004,000A) is FRACTION or FRACTION_SESSION, then only a single Instance of RT PLAN and a single Fraction Group are referenced (i.e., component beams or brachy application setups are not referenced).
| DICOM PS3.3 2024c - Information Object Definitions |
|---|