| DICOM PS3.3 2024c - Information Object Definitions |
|---|
The RT Ion Tolerance Tables Module contains information describing the maximum allowed differences between the planned and measured Attributes for Ion therapy.
Table C.8.8.24-1. RT Ion Tolerance Tables Module Attributes
|
Sequence of ion tolerance tables to be used for delivery of treatment plan. One or more Items shall be included in this Sequence. See Note 1. |
|||
|
Identification number of the Tolerance Table. The value of Tolerance Table Number (300A,0042) shall be unique within the RT Ion Plan in which it is created. |
|||
|
Maximum permitted difference (in degrees) between planned and delivered Gantry Angle. |
|||
|
Maximum permitted difference (in degrees) between planned and delivered Beam Limiting Device Angle. |
|||
|
Type of beam limiting device (collimator). Enumerated Values: |
|||
|
Maximum permitted difference (in mm) between planned and delivered leaf (element) or jaw positions for current beam limiting device (collimator). |
|||
|
Maximum permitted difference (in degrees) between planned and delivered Patient Support Angle. |
|||
|
Maximum permitted difference (in mm) between planned and delivered Table Top Vertical Position. |
|||
|
Maximum permitted difference (in mm) between planned and delivered Table Top Longitudinal Position. |
|||
|
Maximum permitted difference (in mm) between planned and delivered Table Top Lateral Position. |
|||
|
Maximum permitted difference (in degrees) between planned and delivered Table Top Pitch Angle. |
|||
|
Maximum permitted difference (in degrees) between planned and delivered Table Top Roll Angle. |
|||
|
Maximum permitted difference (in mm) between planned and delivered Snout Position. |
|||
|
Maximum permitted difference (in degrees) between planned and delivered Head Fixation Angle (300A,0148). |
|||
|
Maximum permitted difference (in mm) between planned and delivered Chair Head Frame Position (300A,0151). |
|||
|
Maximum permitted difference (in degrees) between planned and delivered Fixation Light Azimuthal Angle (300A,0356). |
|||
|
Maximum permitted difference (in degrees) between planned and delivered Fixation Light Polar Angle (300A,0358). |
| DICOM PS3.3 2024c - Information Object Definitions |
|---|