| DICOM PS3.3 2024c - Information Object Definitions |
|---|
This Module describes the Attributes related to the filtration of X-Rays during the acquisition of an X-Ray image.
Table C.8-35a. X-Ray Filtration Macro Attributes
|
Type of filter(s) inserted into the X-Ray beam (e.g., wedges). |
|||
|
The X-Ray absorbing material used in the filter. May be multi-valued. See Annex P “Correspondence of X-Ray Filter Material Codes and Defined Terms” in PS3.16 for Defined Terms. |
|||
|
The minimum thickness in mm of the X-Ray absorbing material used in the filters. May be multi-valued, with values corresponding to the respective values in Filter Material (0018,7050). |
|||
|
The maximum thickness in mm of the X-Ray absorbing material used in the filters. May be multi-valued, with values corresponding to the respective values in Filter Material (0018,7050). |
|||
|
The minimum length (in mm) of the X-Ray beam path in the Filter Material that is not blocked by collimation. May be multi-valued, with values corresponding to the respective values in Filter Material (0018,7050). |
|||
|
The maximum length (in mm) of the X-Ray beam path in the Filter Material that is not blocked by collimation. May be multi-valued, with values corresponding to the respective values in Filter Material (0018,7050). |
| DICOM PS3.3 2024c - Information Object Definitions |
|---|