| DICOM PS3.3 2024c - Information Object Definitions |
|---|
Table C.8-9 specifies the Attributes of the NM Image Module, which describe Nuclear Medicine Images.
Table C.8-9. NM Image Module Attributes
|
Image identification characteristics. See Section C.8.4.9.1.1 for specialization. |
|||
|
Specifies whether an Image has undergone lossy compression (at a point in its lifetime). Enumerated Values: Once this value has been set to "01" it shall not be reset. Required if Lossy Compression has been performed on the Image. May be present otherwise. |
|||
|
Sum of all gamma events for all frames in the image. See Section C.8.4.9.1.2 for specialization. |
|||
|
Description of how the data collection was stopped. Defined Terms: See Section C.8.4.9.1.3 for specialization. |
|||
|
The height of the patient table in mm. The range and values of this element are determined by the manufacturer. Should not be included if Image Type (0008,0008), Value 3, is |
|||
|
Location of the patient table (or gantry relative to the table) in mm. The range and values of this element are determined by the manufacturer. Should not be included if Image Type (0008,0008), Value 3, is |
|||
|
Elapsed time for data acquisition in msec. Required if Image Type (0008,0008) Value 3 is: See Section C.8.4.9.1.4 for specialization. |
|||
|
Maximum count rate achieved during the acquisition in counts/sec. |
|||
|
Code or description of processing functions applied to the data. |
|||
|
One or more values that indicate which, if any, corrections have been applied to the image. Corrections are applied to all frames in the image. |
|||
|
The type of scan performed. Used only if Image Type (0008,0008), Value 3, contains the value WHOLE BODY. |
|||
|
The speed of the camera motion over the body in mm/sec. Required if Image Type (0008,0008) Value 3 contains the value WHOLE BODY. |
|||
|
Size of the imaged area in the direction of scanning motion, in mm. Required if Image Type (0008,0008) Value 3 contains the value WHOLE BODY. |
|||
|
Include Table 10-7 “General Anatomy Optional Macro Attributes” |
|||
|
The mapping of stored values to associated Real World Values. |
|||
|
>Include Table C.7.6.16-12b “Real World Value Mapping Item Macro Attributes” |
Measurement Units Code Sequence DCID 84 “PET Unit”. |
||
Content Date (0008,0023) and Content Time (0008,0033) are included in the General Image Module, Table C.7-7, whenever the images are temporally related. For this purpose, all NM Images are considered temporally related, so that these elements are included in an NM Image.
Referenced Overlay Sequence (0008,1130) and Referenced Curve Sequence (0008,1145) were previously included in this Module as optional Attributes but have been retired. See PS3.3-2004. Referenced Image Real World Value Mapping Sequence (0040,9094) was previously used in this Module to enclose the Real World Value Mapping Sequence (0040,9096) but has been retired. See PS3.3-2011.
For NM images, Image Type (0008,0008) Value 3 is specified to be Type 1.
Enumerated Values for Value 3:
For NM images, Image Type (0008,0008) Value 1 will be ORIGINAL for all raw data and reconstructed images. DERIVED may be appropriate for some other results images.
For NM images, Image Type (0008,0008) Value 2 will be PRIMARY.
Counts Accumulated (0018,0070) is the total of all gamma events accumulated in all frames of this Image. This Attribute applies to acquisition data, and often does not apply to processed images (DERIVED, SECONDARY).
| DICOM PS3.3 2024c - Information Object Definitions |
|---|