| DICOM PS3.3 2024c - Information Object Definitions |
|---|
Table C.8-6 specifies the Attributes of the NM Image Pixel Module, which describe the pixel data of a NM image.
Table C.8-6. NM Image Pixel Module Attributes
|
Number of samples (color planes) in this image. The value shall be 1. |
|||
|
Specifies the intended interpretation of the pixel data. See Section C.8.4.7.1.1 for further explanation. |
|||
|
Number of bits allocated for each pixel sample. Each sample shall have the same number of bits allocated. |
|||
|
Number of bits stored for each pixel sample. Each sample shall have the same number of bits stored. The value shall be the same as the value in Bits Allocated (0028,0100). |
|||
|
Most significant bit for pixel sample data. Each sample shall have the same high bit. Shall be one less than the value in Bits Stored (0028,0101). |
|||
|
Physical distance in the patient between the center of each pixel, specified by a numeric pair - adjacent row spacing (delimiter) adjacent column spacing, in mm. See Section 10.7.1.3 for further explanation of the value order. |
See Section C.7.6.3.1.2 for definition of these terms.
| DICOM PS3.3 2024c - Information Object Definitions |
|---|