| DICOM PS3.3 2024c - Information Object Definitions |
|---|
Table C.8.32-2 specifies the Attributes of the Parametric Map Image Module.
Table C.8.32-2. Parametric Map Image Module Attributes
|
Image identification characteristics. Value 3 shall be Image Flavor, common Defined Terms for which are specified in Section C.8.16.1.3, CT-specific Defined Terms for which are specified in Section C.8.15.2.1.1.3 and MR-specific Defined Terms for which are specified in Section C.8.13.1.1.1.3. Value 4 shall be Derived Pixel Contrast, common Defined Terms for which are specified in Section C.8.16.1.4, CT-specific Defined Terms for which are specified in Section C.8.15.2.1.1.4 and MR-specific Defined Terms for which are specified in Section C.8.13.1.1.1.4. |
|||
|
Include Table 10-12 “Content Identification Macro Attributes” |
|||
|
Indication of the presence or absence of color information that may be used during rendering. Enumerated Values:
|
|||
|
Number of bits stored for each pixel sample. Required if Pixel Data (7FE0,0010) or Pixel Data Provider URL (0028,7FE0) is present. |
|||
|
Most significant bit for pixel sample data. Required if Pixel Data (7FE0,0010) or Pixel Data Provider URL (0028,7FE0) is present. |
|||
|
Specifies an identity transformation for the Presentation LUT such that the output of all grayscale transformations are defined to be in P-Values. |
|||
|
Specifies whether an Image has undergone lossy compression (at a point in its lifetime), or is derived from lossy compressed images. Enumerated Values: Once this value has been set to "01" it shall not be reset. See Section C.8.32.2.1 and Section C.7.6.1.1.5. |
|||
|
Describes the approximate lossy compression ratio(s) that have been applied to this image. Required if present in the source images or this SOP Instance has been compressed. |
|||
|
A label for the lossy compression method(s) that have been applied to this image. Required if present in the source images or this SOP Instance has been compressed. See Section C.8.32.2.1. |
|||
|
Indicates whether or not image contains sufficient burned in annotation to identify the patient and date the image was acquired. |
|||
|
Indicates whether or not the image contains sufficiently recognizable visual features to allow the image or a reconstruction from a set of images to identify the patient. |
|||
|
Content Qualification Indicator See Section C.8.13.2.1.1 for further explanation. |
|||
|
Palette Color Lookup Table UID. See Section C.7.9.1 for further explanation. Required if Pixel Presentation (0008,9205) equals COLOR_RANGE and the Palette Color Lookup Table Module is not present. |
|||
|
An ICC Profile encoding the transformation of device-dependent color stored pixel values into PCS-Values. When present, defines the color space of the output of the Parametric Map See Section C.11.15.1.1. Required if Pixel Presentation (0008,9205) has a value of COLOR_RANGE. |
|||
|
A label that identifies the well-known color space of the image. Shall be consistent with any ICC Profile (0028,2000) that is also present. See Section C.11.15.1.2. |
|||
If Lossy Image Compression (0028,2110) in any of the source images is "01", the value shall be "01" for the Parametric Map Instance.
The process of derivation of a Parametric Map itself is defined not to be lossy compression, even though it involves loss. If the Parametric Map Instance is encoded using a lossy compression Transfer Syntax, then the value shall be set to "01".
| DICOM PS3.3 2024c - Information Object Definitions |
|---|