Surface Scan Acquisition Type Code Sequence
|
(0080,0001)
|
1
|
Specify the type of acquisition method.
For evaluation of the characteristics or quality of the data.
Only a single Item shall be included in this Sequence.
|
|
>Include Table 8.8-1 “Code Sequence Macro Attributes”
|
BCID 8201 “Surface Scan Acquisition Type”.
|
|
Surface Scan Mode Code Sequence
|
(0080,0002)
|
2
|
Used scan mode.
Zero or more Items shall be included in this Sequence.
|
|
>Include Table 8.8-1 “Code Sequence Macro Attributes”
|
BCID 8202 “Surface Scan Mode Type”.
|
|
Registration Method Code Sequence
|
(0080,0003)
|
1C
|
Method for registering multiple shots.
Required if the IOD stores data that has been derived from multiple shots.
Only a single Item shall be included in this Sequence.
|
|
>Include Table 8.8-1 “Code Sequence Macro Attributes”
|
BCID 8203 “Surface Scan Registration Method Type”.
|
|
Instance Number
|
(0020,0013)
|
1
|
A number that identifies this shot.
Shall start at 1 and monotonically increase by 1 within a single acquisition.
|
|
Acquisition Number
|
(0020,0012)
|
1
|
A number identifying the single continuous gathering of data over a period of time that resulted in this Instance.
|
|
Acquisition DateTime
|
(0008,002A)
|
1
|
The date and time that the acquisition of data that resulted in this Instance started.
|
|
Shot Duration Time
|
(0080,0004)
|
1
|
The time in seconds from the start of this shot until the end of this shot.
|
|
Shot Offset Time
|
(0080,0005)
|
3
|
The time in seconds from the Acquisition DateTime to the start of this shot.
|