Steep Keratometric Axis Sequence
|
(0046,0074)
|
1
|
Steepest meridian as defined by the greatest power of curvature and shortest radius of curvature.
Only a single Item shall be included in this Sequence.
|
|
>Radius of Curvature
|
(0046,0075)
|
1
|
The radius of curvature of the principal meridians of the cornea, measured in mm.
|
|
>Keratometric Power
|
(0046,0076)
|
2
|
The refractive power of the cornea at the principal meridians, measured in diopters.
|
|
>Keratometric Axis
|
(0046,0077)
|
2
|
The meridian where the keratometric radius of curvature or power is measured, in degrees.
|
|
Flat Keratometric Axis Sequence
|
(0046,0080)
|
1
|
Flattest meridian as defined by the least power of curvature and longest radius of curvature.
Only a single Item shall be included in this Sequence.
|
|
>Radius of Curvature
|
(0046,0075)
|
1
|
The radius of curvature of the principal meridians of the cornea, measured in mm.
|
|
>Keratometric Power
|
(0046,0076)
|
2
|
The refractive power of the cornea at the principal meridians, measured in diopters.
|
|
>Keratometric Axis
|
(0046,0077)
|
2
|
The meridian where the keratometric radius of curvature or power is measured, in degrees.
|
|
Keratometry Measurement Type Code Sequence
|
(0022,1096)
|
2
|
Descriptors relevant to keratometry data.
Zero or one Item shall be included in this Sequence.
|
|
>Include Table 8.8-1 “Code Sequence Macro Attributes”
|
DCID 4235 “Keratometry Descriptor”
|
|
Keratometer Index
|
(0022,1033)
|
2
|
The translation factor specific to each keratometer that derives a number for power from the measured radius of curvature of the cornea.
|