| DICOM PS3.3 2024c - Information Object Definitions |
|---|
Table C.8.25.16-2 specifies the Attributes of the Intraocular Lens Calculations Macro.
Table C.8.25.16-2. Intraocular Lens Calculations Macro Attributes
|
Type of refractive surgery a patient has had. Zero or more Items shall be included in this Sequence. Required if the value of Refractive Procedure Occurred (0022,1039) is YES. |
|||
|
The patient's refractive error before any of the refractive surgeries listed in Refractive Surgery Type Code Sequence (0022,1040) were performed. Zero or one Item shall be included in this Sequence. Required if the value of Refractive Procedure Occurred (0022,1039) is YES. |
|||
|
DCID 4240 “Ophthalmic Measurement or Calculation Data Source” |
|||
|
SOP Instance that is relevant to the interpretation of this SOP Instance. Only a single Item shall be included in this Sequence. See Section C.8.25.16.1.1 for further explanation. Required if Source of Corneal Size Data Code Sequence (0022,1036) contains an Item with the value (111784, DCM, "Autorefraction Measurements SOP Instance"). |
|||
|
>>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
|
The value for axial length of the lens in a patient's eye, in mm. |
|||
|
DCID 4240 “Ophthalmic Measurement or Calculation Data Source” |
|||
|
SOP Instance that is relevant to the interpretation of this SOP Instance. Only a single Item shall be included in this Sequence. See Section C.8.25.16.1.1 for further explanation. Required if Source of Lens Thickness Data Code Sequence (0022,1132) contains an Item with the value (111782, DCM, "Axial Measurements SOP Instance"). |
|||
|
>>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
|
DCID 4240 “Ophthalmic Measurement or Calculation Data Source” |
|||
|
SOP Instance that is relevant to the interpretation of this SOP Instance. Only a single Item shall be included in this Sequence. See Section C.8.25.16.1.1 for further explanation. Required if Source of Anterior Chamber Depth Data Code Sequence (0022,1133) contains an Item with the value (111782, DCM, "Axial Measurements SOP Instance"). |
|||
|
>>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
|
Refractive state of the imaged eye at the time of acquisition. |
|||
|
The distance from the corneal vertex of the eye to the back of the corrective lens, in mm. |
|||
|
>>>Include Table 8.8-1 “Code Sequence Macro Attributes” |
DCID 4240 “Ophthalmic Measurement or Calculation Data Source” |
||
|
SOP Instances that are relevant to the interpretation of this SOP Instance. One or more Items shall be included in this Sequence. See Section C.8.25.16.1.1 for further explanation. Required if Source of Refractive Measurements Code Sequence (0022,1135) contains an Item with the value (111783, DCM, "Refractive Measurements SOP Instance"). |
|||
|
>>>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
|
>Include Table C.8.25.16-8 “Cornea Measurement Macro Attributes” |
|||
|
Source of the values of Steep Corneal Axis Sequence (0046,0112) and Flat Corneal Axis Sequence (0046,0113). |
|||
|
DCID 4240 “Ophthalmic Measurement or Calculation Data Source” |
|||
|
SOP Instance that is relevant to the interpretation of this SOP Instance. Only a single Item shall be included in this Sequence. See Section C.8.25.16.1.1 for further explanation. Required if Source of Cornea Measurement Data Code Sequence (0046,0111) contains an Item with the value (111757, DCM, "Keratometry Measurements SOP Instance"). |
|||
|
>>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
|
A free form text description of the of the IOL Formula Code Sequence (0022,1028) (e.g., a reference to the mathematical equation). |
|||
|
Include Table C.8.25.16-4 “IOL Ophthalmic Axial Length Macro Attributes” |
|||
|
The astigmatism that is expected to be induced by corneal incisions during cataract surgery. |
|||
| DICOM PS3.3 2024c - Information Object Definitions |
|---|