| DICOM PS3.3 2024c - Information Object Definitions |
|---|
Table C.8.25.14-5 specifies the Attributes of the Ophthalmic Axial Measurements Selected Macro. These are the Attributes of the Ophthalmic Axial Length (0022,1019) selected for use in IOL calculations.
Table C.8.25.14-5. Ophthalmic Axial Measurements Selected Macro Attributes
|
Information related to the selected axial length measurement of the patient's eye when acquired on an ultrasound device. Only a single Item shall be included in this Sequence. Required if Ophthalmic Axial Measurements Device Type (0022,1009) is ULTRASOUND. |
|||
|
Identifies whether measuring the total axial length of the patient's eye or a segment of the eye for which a discrete measurement was obtained. |
|||
|
The axial length measurement, in mm. The type of measurement is specified in the Ophthalmic Axial Length Measurements Type (0022,1010). See Section C.8.25.14.1.1.4 for further explanation. |
|||
|
Method used to derive the value recorded in Ophthalmic Axial Length (0022,1019). |
|||
|
>>Include Table 8.8-1 “Code Sequence Macro Attributes”. |
|||
|
>Referenced Ophthalmic Axial Length Measurement QC Image Sequence |
References to the quality control image associated with this measurement. Only a single Item shall be included in this Sequence. See Section C.8.25.14.1.1.6 for further explanation. |
||
|
>>Include Table C.8.25.16-6 “Ophthalmic Axial Measurements Quality Image SOP Instance Reference Macro Attributes” |
|||
|
Information about the quality metric applied to Ophthalmic Axial Length (0022,1019). |
|||
|
>>Include Table C.8.25.14-6 “Ophthalmic Axial Length Quality Metric Macro Attributes” |
|||
|
Segmental axial length measurements of a patient's eye that are summed to derive a total axial length for the eye. One or more Items shall be included in this Sequence. Required if the value of Ophthalmic Axial Length Measurements Type (0022,1010) is present and is LENGTH SUMMATION. May be present otherwise. |
|||
|
The axial length measurement, in mm. The type of measurement is specified in the Ophthalmic Axial Length Measurements Type (0022,1010). See Section C.8.25.14.1.1.3 and Section C.8.25.14.1.1.4 for further explanation. |
|||
|
>>Ophthalmic Axial Length Measurements Segment Name Code Sequence |
Name of the segment measured. See Section C.8.25.14.1.1.4 for further explanation. |
||
|
>>>Include Table 8.8-1 “Code Sequence Macro Attributes” |
DCID 4233 “Ophthalmic Axial Length Measurements Segment Name” |
||
|
Information related to the selected axial length measurement(s) of the patient's eye when acquired on an optical device. One or more Items shall be included in this Sequence. Required if Ophthalmic Axial Measurements Device Type (0022,1009) is OPTICAL. |
|||
|
Identifies whether measuring the total axial length of the patient's eye or a segment of the eye for which a discrete measurement was obtained. |
|||
|
Total axial length measurement selected for the patient's eye. Only a single Item shall be included in this Sequence. Required if the value of Ophthalmic Axial Length Measurements Type (0022,1010) is present and is either TOTAL LENGTH or LENGTH SUMMATION. May be present otherwise |
|||
|
The axial length measurement, in mm. The type of measurement is specified in the Ophthalmic Axial Length Measurements Type (0022,1010). See Section C.8.25.14.1.1.3 and Section C.8.25.14.1.1.4 for further explanation. |
|||
|
>>Referenced Ophthalmic Axial Length Measurement QC Image Sequence |
Reference to the quality control image associated with this measurement. Only a single Item shall be included in this Sequence. See Section C.8.25.14.1.1.6 for further explanation. |
||
|
>>>Include Table C.8.25.16-6 “Ophthalmic Axial Measurements Quality Image SOP Instance Reference Macro Attributes” |
|||
|
Information about the quality metric applied to Ophthalmic Axial Length (0022,1019). |
|||
|
>>>Include Table C.8.25.14-6 “Ophthalmic Axial Length Quality Metric Macro Attributes” |
|||
|
Segmental axial length measurement(s) selected for the patient's eye. One or more Items shall be included in this Sequence. Required if the value of Ophthalmic Axial Length Measurements Type (0022,1010) is present and is either SEGMENTAL LENGTH or LENGTH SUMMATION. May be present otherwise. |
|||
|
>>Ophthalmic Axial Length Measurements Segment Name Code Sequence |
Name of the segment measured. See Section C.8.25.14.1.1.4 for further explanation. |
||
|
>>>Include Table 8.8-1 “Code Sequence Macro Attributes” |
DCID 4233 “Ophthalmic Axial Length Measurements Segment Name” |
||
|
The axial length measurement, in mm. The type of measurement is specified in the Ophthalmic Axial Length Measurements Type (0022,1010). See Section C.8.25.14.1.1.3 and Section C.8.25.14.1.1.4 for further explanation. |
|||
|
>>Referenced Ophthalmic Axial Length Measurement QC Image Sequence |
Reference to the quality control image associated with this measurement. Only a single Item is permitted in this Sequence. See Section C.8.25.14.1.1.6 for further explanation. |
||
|
>>>Include Table C.8.25.16-6 “Ophthalmic Axial Measurements Quality Image SOP Instance Reference Macro Attributes” |
|||
|
Information about the quality metric applied to Ophthalmic Axial Length (0022,1019). |
|||
|
>>>Include Table C.8.25.14-6 “Ophthalmic Axial Length Quality Metric Macro Attributes” |
|||
| DICOM PS3.3 2024c - Information Object Definitions |
|---|