| DICOM PS3.3 2024c - Information Object Definitions |
|---|
Table C.8.25.14-3 specifies the Attributes of the Ophthalmic Axial Length Segmental Measurements Macro.
Table C.8.25.14-3. Ophthalmic Axial Length Segmental Measurements Macro Attributes
|
The axial length measurement, in mm. The type of measurement is specified in the Ophthalmic Axial Length Measurements Type (0022,1010). See Section C.8.25.14.1.1.3 and Section C.8.25.14.1.1.4 for further explanation. |
|||
|
Whether or not the clinician intervened to modify the output of the device. For example by forcing it to select a different peak in the display. |
|||
|
Ophthalmic Axial Length Measurements Segment Name Code Sequence |
The name of the segment measured. See Section C.8.25.14.1.1.4 for further explanation. |
||
|
DCID 4233 “Ophthalmic Axial Length Measurements Segment Name” |
|||
|
Include Table C.8.25.14-4 “Ophthalmic Axial Measurements Related Information Macro Attributes” |
|||
| DICOM PS3.3 2024c - Information Object Definitions |
|---|