DICOM PS3.3 2024c - Information Object Definitions
C.8.22.5.3 PET Detector Motion Details Macro

Table C.8.22-12 specifies the Attributes of the PET Detector Motion Details Macro.

Table C.8.22-12. PET Detector Motion Details Macro Attributes

Attribute Name

Tag

Type

Attribute Description

PET Detector Motion Details Sequence

(0018,9733)

1

Contains the Attributes defining the details of the motion of the Detector.

Only a single Item shall be included in this Sequence.

>Rotation Direction

(0018,1140)

1

Direction of rotation of the detector about the gantry, as viewed while facing the gantry where the table enters the gantry.

Enumerated Values:

CW

clockwise

CC

counter clockwise

>Revolution Time

(0018,9305)

1

The time in seconds of a complete revolution of the detector around the gantry orbit.


DICOM PS3.3 2024c - Information Object Definitions