| DICOM PS3.3 2024c - Information Object Definitions |
|---|
The following sections contain Macros specific to the Segmentation IOD.
Table C.8.20-4 specifies the Attributes of the Segment Description Macro.
Table C.8.20-4. Segment Description Macro Attributes
|
Identification number of the segment. The value of Segment Number (0062,0004) shall be unique within the Segmentation Instance in which it is created. See Section C.8.20.2.4. |
|||
|
User-defined label identifying this segment. This may be the same as Code Meaning (0008,0104) of Segmented Property Type Code Sequence (0062,000F). |
|||
|
Include Table 10-7b “Multiple Site General Anatomy Optional Macro Attributes” |
May not be necessary if the anatomy is implicit in the Segmented Property Type Code Sequence. More than one Item in Anatomic Region Sequence (0008,2218) may be used when a region of interest spans multiple anatomical locations and there is not a single pre-coordinated code describing the combination of locations. There is no requirement that the multiple locations be contiguous. |
||
|
Sequence defining the general category of the property the segment represents. |
|||
|
Sequence defining the specific property the segment represents. Note"Property" is used in the sense of meaning "what the segmented voxels represent", whether it be a physical or biological object, be real or conceptual, having spatial, temporal or functional extent or not. I.e., it is what the segment "is" (as opposed to some feature, attribute, quality, or characteristic of it, like color or shape or size). |
|||
|
Sequence defining the modifier of the property type of this segment. |
|||
|
A text label used for tracking a finding or feature, potentially across multiple reporting objects, over time. This label shall be unique within the domain in which it is used. Required if Tracking UID (0062,0021) is present. Note
|
|||
|
A unique identifier used for tracking a finding or feature, potentially across multiple reporting objects, over time. Required if Tracking ID (0062,0020) is present. Note
|
|||
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes”. |
|||
|
The value of ROI Number (3006,0022) in the referenced SOP Instance that identifies the ROI that is the origin of the Segment information. Required if Referenced SOP Class UID (0008,1150) is RT Structure Set Storage ("1.2.840.10008.5.1.4.1.1.481.3"). |
|||
| DICOM PS3.3 2024c - Information Object Definitions |
|---|