| DICOM PS3.3 2024c - Information Object Definitions |
|---|
Table C.8.19.6-2 specifies the Attributes of the X-Ray Field of View Macro.
Table C.8.19.6-2. X-Ray Field of View Macro Attributes
|
Shape of the Field of View, that is the image pixels stored in Pixel Data (7FE0,0010). |
|||
|
Dimensions in mm of the Field of View, that is the image pixels stored in Pixel Data (7FE0,0010). If Field of View Shape (0018,1147) is: RECTANGLE: row dimension followed by column. HEXAGONAL: diameter of the circle circumscribing the hexagon. |
|||
|
Offset of the TLHC of a rectangle circumscribing the Field of View, i.e., the image pixels stored in Pixel Data (7FE0,0010) before rotation or flipping, from the TLHC of the physical detector area measured in physical detector pixels as a row offset followed by a column offset. See Section C.8.11.4.1.1 for further explanation. Required if X-Ray Receptor Type (0018,9420) is present and equals DIGITAL_DETECTOR. May be present otherwise. |
|||
|
Clockwise rotation in degrees of Field of View, i.e., the image pixels stored in Pixel Data (7FE0,0010), relative to the physical detector. See Section C.8.11.4.1.1 for further explanation. |
|||
|
Whether or not a horizontal flip has been applied to the Field of View, i.e., the image pixels stored in Pixel Data (7FE0,0010), after rotation relative to the physical detector as described in Field of View Rotation (0018,7032). See Section C.8.11.4.1.1 for further explanation. |
|||
|
Manufacturer defined description of the field of view selected during acquisition. |
| DICOM PS3.3 2024c - Information Object Definitions |
|---|